Defining the <strong>System</strong> Setup: Define the <strong>ECG</strong> Setup<br />
Item<br />
Description<br />
Item<br />
Description<br />
Rhythm leads<br />
Autorhythm<br />
Choose a lead option for each group to determine the<br />
rhythm leads that print when you select the rhythm<br />
key in an application. The six defined groups make up<br />
the display list when you select Leads in the <strong>Resting</strong>,<br />
Pediatric, 15 Lead, or Vector Loops Application.<br />
n Select 3 leads to define which three leads in a three<br />
lead Rhythm report print.<br />
n Select 6 leads to define which six leads in a six lead<br />
Rhythm report print.<br />
n Select All leads to display and print 10 seconds of<br />
data for 12 (or 15) leads.<br />
n Select Lead Check to display and print real time<br />
data for each of the 12 (or 15) leads.<br />
n Select Lead Placement to display and print real<br />
time data for each of the 12 (or 15) leads and the a<br />
chest electrode placement.<br />
Select the group of Rhythm leads printed in the<br />
Autorhythm report.<br />
RMR/CGR/extra<br />
rhythm leads<br />
Swedish format<br />
rhythm leads<br />
Select the rhythm lead(s) you want printed in the RMR<br />
and CGR reports.<br />
áWhen you change a rhythm lead, the new lead<br />
appears on all reports displaying that lead. For<br />
example, if you select V5 for RMR/CGR/extra<br />
rhythm lead 1, then the V5 waveform appears on<br />
all reports that include RMR/CGR/extra rhythm<br />
lead 1.<br />
Select the rhythm lead(s) you want printed in the<br />
Swedish format reports.<br />
áWhen you change a rhythm lead, the new lead<br />
appears on all reports displaying that lead. For<br />
example, if you select V5 for the Swedish format<br />
rhythm lead 1, then the V5 waveform appears on<br />
all reports that include Swedish format rhythm<br />
lead 1.<br />
<strong>MAC</strong> <strong>5000</strong> <strong>System</strong> — 2000657-033, Revision A 14-21