Define the <strong>ECG</strong> Setup<br />
<strong>ECG</strong> Acquisition/<strong>Analysis</strong><br />
Select this function to define the <strong>ECG</strong> acquisition and<br />
analysis parameters.<br />
<strong>System</strong> Setup<br />
<strong>ECG</strong><br />
Item<br />
Baseline roll filter<br />
Pre-acquisition<br />
<strong>ECG</strong><br />
<strong>ECG</strong> Acquisition/<strong>Analysis</strong><br />
Description<br />
168A, 057A<br />
Use this filter to remove baseline sway.<br />
áThe higher the setting, the more the filter smooths<br />
out a wandering baseline. This filter does NOT<br />
distort the ST segment displayed on the <strong>ECG</strong><br />
reports.<br />
Select Yes to begin acquiring <strong>ECG</strong> data as soon as<br />
the leadwires are connected to a patient.The system<br />
does not wait until the user presses ecg before it<br />
starts acquiring <strong>ECG</strong> data. The latest 10 seconds of<br />
<strong>ECG</strong> data is ready for analysis when Pre-acquisition is<br />
turned on.<br />
Item<br />
Disable auto gain<br />
check<br />
Disable lead off<br />
check<br />
Baseline wander<br />
warning<br />
Muscle tremor<br />
warning<br />
AC filter<br />
AC noise level<br />
warning<br />
Screening<br />
criteria<br />
Defining the <strong>System</strong> Setup: Define the <strong>ECG</strong> Setup<br />
Select No, to display a prompt after the user presses<br />
ecg if the gain of the recorded <strong>ECG</strong> data is either too<br />
high or too low. The user can adjust the gain.<br />
Select No to display a screen message when the<br />
system detects a disconnected leadwire.<br />
Select Yes to display a screen message when the<br />
system detects a wandering baseline.<br />
Select Yes to display a screen message when the<br />
system detects muscle tremor.<br />
Use this filter to remove AC line artifact.<br />
Select Yes to program the system to check for<br />
alternating current interference when recording an<br />
<strong>ECG</strong>.<br />
Select Yes to prevent specific 12SL analysis<br />
statements from appearing on <strong>ECG</strong> reports.<br />
See Appendix C to identify these statements.<br />
Suppress<br />
NORMAL<br />
statement<br />
Description<br />
Select Yes to prevent the Normal <strong>ECG</strong> 12SL analysis<br />
statement from appearing on printed, stored, and<br />
transmitted <strong>ECG</strong> reports.<br />
<strong>MAC</strong> <strong>5000</strong> <strong>System</strong> — 2000657-033, Revision A 14-17