Define the <strong>ECG</strong> Setup<br />

<strong>ECG</strong> Acquisition/<strong>Analysis</strong><br />

Select this function to define the <strong>ECG</strong> acquisition and<br />

analysis parameters.<br />

<strong>System</strong> Setup<br />

<strong>ECG</strong><br />

Item<br />

Baseline roll filter<br />

Pre-acquisition<br />

<strong>ECG</strong><br />

<strong>ECG</strong> Acquisition/<strong>Analysis</strong><br />

Description<br />

168A, 057A<br />

Use this filter to remove baseline sway.<br />

áThe higher the setting, the more the filter smooths<br />

out a wandering baseline. This filter does NOT<br />

distort the ST segment displayed on the <strong>ECG</strong><br />

reports.<br />

Select Yes to begin acquiring <strong>ECG</strong> data as soon as<br />

the leadwires are connected to a patient.The system<br />

does not wait until the user presses ecg before it<br />

starts acquiring <strong>ECG</strong> data. The latest 10 seconds of<br />

<strong>ECG</strong> data is ready for analysis when Pre-acquisition is<br />

turned on.<br />

Item<br />

Disable auto gain<br />

check<br />

Disable lead off<br />

check<br />

Baseline wander<br />

warning<br />

Muscle tremor<br />

warning<br />

AC filter<br />

AC noise level<br />

warning<br />

Screening<br />

criteria<br />

Defining the <strong>System</strong> Setup: Define the <strong>ECG</strong> Setup<br />

Select No, to display a prompt after the user presses<br />

ecg if the gain of the recorded <strong>ECG</strong> data is either too<br />

high or too low. The user can adjust the gain.<br />

Select No to display a screen message when the<br />

system detects a disconnected leadwire.<br />

Select Yes to display a screen message when the<br />

system detects a wandering baseline.<br />

Select Yes to display a screen message when the<br />

system detects muscle tremor.<br />

Use this filter to remove AC line artifact.<br />

Select Yes to program the system to check for<br />

alternating current interference when recording an<br />

<strong>ECG</strong>.<br />

Select Yes to prevent specific 12SL analysis<br />

statements from appearing on <strong>ECG</strong> reports.<br />

See Appendix C to identify these statements.<br />

Suppress<br />

NORMAL<br />

statement<br />

Description<br />

Select Yes to prevent the Normal <strong>ECG</strong> 12SL analysis<br />

statement from appearing on printed, stored, and<br />

transmitted <strong>ECG</strong> reports.<br />

<strong>MAC</strong> <strong>5000</strong> <strong>System</strong> — 2000657-033, Revision A 14-17