Chapter 5<br />
Patient Preparation<br />
Preparing Patients for Exercise Stress Testing<br />
Apply electrodes and lead wires before entering patient data into the Eclipse.<br />
This allows time for the electrodes to adhere and improves conductivity.<br />
1. Place limb leads where movement is not a factor (see Figure<br />
5-7).<br />
2. Vigorous skin preparation is essential for preventing artifact<br />
during exercise stress testing:<br />
4 Shave all hair from sensor sites<br />
4 Cleanse sensor sites with alcohol<br />
4 Use a Skin Rasp. Applying moderate pressure, stroke the<br />
skin 2 or 3 times at each site with the rough side of the rasp.<br />
3. As you apply sensors, press down on the outside foam area Þrst<br />
to avoid squeezing the gel out.<br />
4. After applying the sensors, follow the instructions on page 5-6,<br />
ÒConnecting Cables and Lead WiresÓ.<br />
5-10 Eclipse <strong>Operating</strong> <strong>Instructions</strong>