Chapter 4<br />

System Setup and User Setup<br />

MISCELLANEOUS SETTINGS<br />

Baseline Filter<br />

Pacer Enhancement<br />

0.05 Hz<br />

STABLE<br />

Sets the low frequency response.<br />

Selecting 0.05 Hz provides the greatest low-frequency response with no delay and<br />

should be used when no baseline drift is present.<br />

Selecting STABLE (Baseline Stabilization Filter) will effectively suppress most<br />

baseline wander interference while providing an accurate ECG reproduction (no<br />

distortion of the ST segment.) Use of the Baseline Stabilization Filter will result in an<br />

approximately 1.5 second delay in the displayed ECG. This filter is applied to the<br />

printed report, displayed ECG, and the data analyzed.<br />

NOTE: The American Heart Association’s 1990 Recommendations<br />

(Recommendations for Standardization and Specifications in Automated<br />

Electrocardiography: Bandwidth and Signal Processing) pertaining to low-frequency<br />

response in electrocardiography are met or exceeded by both the 0.05 Hz and<br />

STABLE selections.<br />

ON<br />

OFF<br />

Enables pacemaker enhancement. This feature makes pacemaker signals show up<br />

as prominent spikes on the display and on printouts. An enhanced pacer is printed<br />

as a spike that is at least 10 mm tall and of positive polarity.<br />

Rhythm Mode Pages [Numeric. Range = 1-10]<br />

Sets the number of pages printed during an Auto Rhythm.<br />

Bradycardia Limit [Numeric. Range = 41-69]<br />

Reports for adult patients with heart rates below this limit are labelled<br />

“BRADYCARDIA.” For more information on heart rate limits refer to the Physician’s<br />

Guide.<br />

NOTE: This feature is available on interpretive units only.<br />

Tachycardia Limit [Numeric. Range = 81-129]<br />

Reports for adult patients with heart rates above this limit are labelled<br />

“TACHYCARDIA.” For more information on heart rate limits refer to the Physician’s<br />

Guide.<br />

NOTE: This feature is available on interpretive units only.<br />

Password<br />

[Up to 15 alphanumeric characters]<br />

NOTE: The password field will not appear in the MISCELLANEOUS menu if the<br />

password is set to OFF in SYSTEM SETUP.<br />

NOTE: Write down the password and keep it in a secure place. You will be able to<br />

change or remove the password protection only after you have entered the correct<br />

password.<br />

Determines the password required to access the USER SETUP menus and the<br />

SYSTEM SETUP menu. You can set a different password for both User 1 and User<br />

2. This field is enabled in the SYSTEM SETUP menu [see “Password” on pg. 4-5].<br />

4-12 Eclipse <strong>Operating</strong> <strong>Instructions</strong>