ConÞguring the USER SETUP Menus<br />

Analysis-copies<br />

ON<br />

OFF<br />

Determines whether analysis statements appear on copies of Auto ECG reports.<br />

Number Of Copies [Numeric. Range = 0-5]<br />

Sets the number of complete report copies to print in addition to the automatic<br />

printout.<br />

Median Complex Page<br />

Print Mode<br />

Save Mode<br />

Wait for good data<br />

ON<br />

OFF<br />

Enables Median Complex printout as part of every Auto ECG report.<br />

AUTO<br />

MANUAL<br />

OFF<br />

Determines how ECG reports are printed. If AUTO is selected, then reports<br />

automatically print out prior to saving. If OFF is selected, no report is printed prior to<br />

saving. If MANUAL is selected, then prior to saving the unit will display a message<br />

asking whether to print the record.<br />

AUTO<br />

AUTO SAVE AND SEND<br />

MANUAL<br />

Determines how Auto ECG reports are saved. If AUTO is selected, reports will be<br />

automatically saved after printing. If MANUAL save is selected, the PREVIEW<br />

screen is displayed at the end of every Auto ECG sequence. You may save the report<br />

at this time. If AUTO SAVE AND SEND is selected, then the report is automatically<br />

saved; the report is also sent automatically to an ECG Management System, such<br />

as the PYRAMIS System or the FAA System.<br />

ON<br />

OFF<br />

When ON, Auto ECG reports are delayed until ten seconds of low noise waveform<br />

have been collected (all electrodes must be connected). An immediate report can be<br />

forced by pressing ECG a second time. In this case, the report will state “Warning:<br />

Data quality may affect computer interpretation.”<br />

When OFF, the Auto ECG report starts once at least 10 seconds of ECG waveform<br />

has been collected, regardless of the quality of the collected waveform data. No<br />

warning message is printed in this case.<br />

NOTE: If any electrodes are disconnected the report will show a flat baseline for<br />

those leads. (“Overload” conditions will also produce a flat baseline.) Leads with<br />

disconnected electrodes will not be used in the ECG analysis.<br />

CUSTOM LEAD 1 & CUSTOM LEAD 2<br />

These two menus each format a group of leads. These groups can later be<br />

selected by pressing the Ò5Ó or Ò6Ó machine control keys (see ÒThe KeyboardÓ<br />

on pg. 3-1).<br />

Custom Lead Group 1 is also used for the rhythm page of an Auto ECG report.<br />

Only Standard or Cabrera conÞgurations are acceptable for this purpose. Do<br />

not select Frank or Nehb if you want a rhythm page as part of an Auto ECG.<br />

Eclipse <strong>Operating</strong> <strong>Instructions</strong> 4-9