ConÞguring the USER SETUP Menus<br />
Analysis-copies<br />
ON<br />
OFF<br />
Determines whether analysis statements appear on copies of Auto ECG reports.<br />
Number Of Copies [Numeric. Range = 0-5]<br />
Sets the number of complete report copies to print in addition to the automatic<br />
printout.<br />
Median Complex Page<br />
Print Mode<br />
Save Mode<br />
Wait for good data<br />
ON<br />
OFF<br />
Enables Median Complex printout as part of every Auto ECG report.<br />
AUTO<br />
MANUAL<br />
OFF<br />
Determines how ECG reports are printed. If AUTO is selected, then reports<br />
automatically print out prior to saving. If OFF is selected, no report is printed prior to<br />
saving. If MANUAL is selected, then prior to saving the unit will display a message<br />
asking whether to print the record.<br />
AUTO<br />
AUTO SAVE AND SEND<br />
MANUAL<br />
Determines how Auto ECG reports are saved. If AUTO is selected, reports will be<br />
automatically saved after printing. If MANUAL save is selected, the PREVIEW<br />
screen is displayed at the end of every Auto ECG sequence. You may save the report<br />
at this time. If AUTO SAVE AND SEND is selected, then the report is automatically<br />
saved; the report is also sent automatically to an ECG Management System, such<br />
as the PYRAMIS System or the FAA System.<br />
ON<br />
OFF<br />
When ON, Auto ECG reports are delayed until ten seconds of low noise waveform<br />
have been collected (all electrodes must be connected). An immediate report can be<br />
forced by pressing ECG a second time. In this case, the report will state “Warning:<br />
Data quality may affect computer interpretation.”<br />
When OFF, the Auto ECG report starts once at least 10 seconds of ECG waveform<br />
has been collected, regardless of the quality of the collected waveform data. No<br />
warning message is printed in this case.<br />
NOTE: If any electrodes are disconnected the report will show a flat baseline for<br />
those leads. (“Overload” conditions will also produce a flat baseline.) Leads with<br />
disconnected electrodes will not be used in the ECG analysis.<br />
CUSTOM LEAD 1 & CUSTOM LEAD 2<br />
These two menus each format a group of leads. These groups can later be<br />
selected by pressing the Ò5Ó or Ò6Ó machine control keys (see ÒThe KeyboardÓ<br />
on pg. 3-1).<br />
Custom Lead Group 1 is also used for the rhythm page of an Auto ECG report.<br />
Only Standard or Cabrera conÞgurations are acceptable for this purpose. Do<br />
not select Frank or Nehb if you want a rhythm page as part of an Auto ECG.<br />
Eclipse <strong>Operating</strong> <strong>Instructions</strong> 4-9