The PREVIEW Screen<br />
(A) PATIENT ID<br />
IdentiÞcation number of the current patient. On the Eclipse Plus, this Þeld also<br />
displays the patientÕs name.<br />
(B) RELATIVE GAIN<br />
Tracks the amplitudes of the displayed waveforms. This corresponds to the<br />
Recorder Gain setting as follows:<br />
Recorder Gain<br />
Relative Gain<br />
5 mm/mV x1/2<br />
10 mm/mV x1<br />
20 mm/mV x2<br />
L=10, C=5 mm/mV L=x1, C=x1/2<br />
L=20, C=10 mm/mV<br />
L=x2, C=x1<br />
(C) HEART RATE<br />
Displays the patientÕs heart rate.<br />
(D) FUNCTIONS LIST<br />
Displays available functions and menus. Displays space available (Eclipse<br />
PLUS only), for example, ÒAvailable: 60 ECGÓ.<br />
(E) RECORDER SPEED, GAIN AND FREQUENCY RESPONSE<br />
Displays settings for current ECG reports.<br />
(F) BATTERY STATUS GAUGE (ECLIPSE PLUS)<br />
Tracks battery charge level.<br />
(G) POWER STATUS<br />
Indicates the power source, AC line power or battery.<br />
(H) SENSOR STATUS<br />
Indicates signal quality.<br />
If all sensors are producing good signals, the display reads:<br />
SENSORS OK<br />
If one or more sensors produce a poor signal, the labels for the affected sensors<br />
are highlighted and the problem is indicated by one of the following messages:<br />
FAIL<br />
DRIFT<br />
NOISE<br />
NOTE: To ensure accurate sensor status indication, the RL<br />
electrode must be connected.<br />
NOTE: Refer to Chapter 10, ÒTroubleshootingÓ for<br />
information on correcting signal problems.<br />
Eclipse <strong>Operating</strong> <strong>Instructions</strong> 3-5