Technical SpeciÞcations (Eclipse Plus)<br />

Technical Specifications (Eclipse Plus)<br />

Dimensions (with cart):<br />

Weight (with cart):<br />

Power Requirements:<br />

mains power requirement<br />

battery operation<br />

Fuses:<br />

F1 and F2 mains<br />

Environmental:<br />

operating temperature<br />

storage temperature<br />

relative humidity<br />

Acquisition:<br />

atmospheric pressure<br />

lead selection<br />

interpretation<br />

modes<br />

frequency response<br />

input impedance<br />

electrode offset tolerance<br />

a/d conversion<br />

Storage Resolution:<br />

Artifact Filter Response:<br />

Display:<br />

Printout:<br />

paper type<br />

chart speeds<br />

gain<br />

printout formats<br />

printout device<br />

paper dimension<br />

43.5cm x 65.1cm x 125.7cm<br />

(17.125" x 25.625" x 49.5")<br />

39.5 kg (87 lbs)<br />

115/230 V AC ±10%, 0.80/0.40 A, 50/60 Hz<br />

16.8 VDC NiCd battery pack<br />

for 115 V, 1.0 A 250 V type T<br />

for 230 V, 0.500 A 250 V type T<br />

10° C to 40° C<br />

-20° C to 55° C<br />

25% to 95% non-condensing<br />

7 x 10 4 to 10.6 x 10 4 Pa<br />

I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6<br />

Supports Frank X,Y,Z; Nehb D,A,J; and Alternate<br />

chest lead<br />

diagnosis, measurements, reasons statements<br />

auto, rhythm, manual<br />

meets or exceeds ANSI/AAMI EC11-1991 standard<br />

meets or exceeds ANSI/AAMI EC11-1991 standard<br />

±300 mV<br />

5µV LSB<br />

500 samples/sec, 5µV resolution<br />

40 Hz, -3dB<br />

640 x 480 pixel Active-Matrix, Color,<br />

Liquid Crystal Display (LCD)<br />

thermal sensitive<br />

10, 25, 50 mm/sec (meets or exceeds ANSI/AAMI<br />

EC11-1991 standard)<br />

5, 10, 20 mm/mV (meets or exceeds ANSI/AAMI<br />

EC11-1991 standard)<br />

3, 4, 6 or 12 channels; additional rhythm formats<br />

216mm thermal dot array<br />

8.5” x 11” Z-fold or<br />

210mm x 300mm (A4) Z-fold<br />

Eclipse <strong>Operating</strong> <strong>Instructions</strong> 11-9