Chapter 11<br />
Maintenance and Service<br />
These units have been tested for electromagnetic compatibility in accordance<br />
with IEC 601-1-2. The failure criterion for the device is, Òundetectable<br />
interference with the ECG trace which could result in misdiagnosis.Ó While the<br />
device passes the relevant standards, it may exhibit evidence of interference<br />
when subjected to electrostatic discharges, high voltage transients or high<br />
voltage surges, as deÞned in IEC 801-2, IEC 801-4 and IEC 801-5. The<br />
interference from a single event will demonstrate as a sharp noise spike on the<br />
ECG trace. The clinician will not confuse such a noise spike with the heart beat<br />
waveforms and there is no hazard of misdiagnosis. In the unlikely situation<br />
that the equipment is placed in an environment where such interference events<br />
are common, either the equipment or the interference source may be moved.<br />
Battery pack<br />
Under normal conditions, a fully charged battery pack provides adequate<br />
power to print a minimum of 200 pages of data at paper speed of 25 mm/sec;<br />
or a minimum of 20 minutes of continuous acquisition and printing.<br />
A fully discharged battery pack will completely recharge in approximately 5<br />
hours.<br />
Heart rate detection<br />
The Eclipse heart rate meter is speciÞed to function in the range of 30 to 250<br />
bpm with a tolerance of ±10% or ±5 bpm, whichever is greater. Question marks<br />
(???) appear on the printout if the heart rate is out of the speciÞed range.<br />
11-6 Eclipse <strong>Operating</strong> <strong>Instructions</strong>