Chapter 10<br />

Troubleshooting<br />

To reduce electrical interference:<br />

4 Keep the power cord away from the patient and patient<br />

cable.<br />

4 Connect the unit to a properly grounded wall outlet.<br />

4 Arrange the patient cable leads together, closely following<br />

the body contour.<br />

4 Check the line Þlter setting in the SYSTEM SETUP menu.<br />

For more information, see ÒConÞguring SYSTEM SETUP<br />

MenuÓ on pg. 4-1.<br />

4 Ensure that Diathermy or X-ray equipment in adjacent<br />

rooms is not operating. Other electrical equipment<br />

including electric beds, televisions, and lighting Þxtures<br />

may also generate interference (even when not in use).<br />

4 Try moving the patient to another place in the room.<br />

Sometimes, electrical wiring in walls and ceilings causes<br />

interference.<br />

4 Operate the Eclipse from battery power.<br />

The Defibrillation Recovery filter<br />

NOTE: The DeÞbrillation Recovery Þlter should be used only<br />

for enhanced recovery time after deÞbrillation to conÞrm the<br />

presence of cardiac activity. While this Þlter is enabled,<br />

waveforms printed are not of diagnostic quality. Care must be<br />

taken by the operator to not misinterpret the patientÕs condition<br />

during an ÒOVERLOADÓ condition.<br />

The Eclipse provides a DeÞbrillation Recovery Þlter for faster recovery time<br />

from an ÒOVERLOADÓ condition caused by deÞbrillation (see<br />

ÒÒOVERLOADÓ messageÓ on pg. 10-4).<br />

This Þlter is enabled by pressing the ÒBÓ key. Do this only after deÞbrillation<br />

and only if the trace does not recover. When the Þlter is enabled, the Sensor<br />

Status message in the PREVIEW screen displays ÒFAILÓ and ÒOVERLOADÓ is<br />

printed on the report. After approximately 30 seconds, the Þlter clears and the<br />

Eclipse returns to normal operation.<br />

10-6 Eclipse <strong>Operating</strong> <strong>Instructions</strong>