Chapter 10<br />

Troubleshooting<br />

PROBLEM<br />

Problems using full format paper.<br />

REDUCED PERFORMANCE MODE<br />

Paper speed erratic. ECG printouts have<br />

unexpected breaks in the waveforms at<br />

random intervals. Breaks are<br />

accompanied by vertical dashed lines,<br />

header/footer information and lead<br />

designators.<br />

Report is printing ÒSTATÓ in the Patient<br />

ID Þeld of the report.<br />

PROBABLE CAUSE<br />

- Load paper (see ÒLoading Recording<br />

PaperÓ on pg. 2-6)<br />

- Ensure the paper is at top-of-form by<br />

pressing the ÒPÓ key and waiting until<br />

the paper stops. If ÒPaper OUT:<br />

Reload and continue?Ó message<br />

appears, select OK.<br />

- Press ÒMANÓ. When printing starts,<br />

press ÒSTOPÓ.<br />

- Again, press ÒMANÓ. When printing<br />

starts, press ÒSTOPÓ.<br />

- 50 mm/s paper speed selected.<br />

Reduce paper speed to 25 mm/s.<br />

- Artifact Þlter is set to 40 Hz (on).<br />

Change setting to 150 Hz.<br />

- Current Custom Lead setting is Frank.<br />

Use another Custom Lead setting or<br />

reduce paper speed and turn off<br />

Artifact Filter.<br />

- No ID number was entered for the<br />

patient.<br />

Recognizing and Reducing ECG Artifacts<br />

“OVERLOAD” message<br />

DeÞbrillating the patient will interfere with the Eclipse sensors. This may<br />

result in loss of the trace or erratic trace deßections. An ÒOVERLOADÓ<br />

message alerts you that waveforms may not be accurate. To speed recovery<br />

time after deÞbrillation press the ÒBÓ key to activate the DeÞbrillation<br />

Recovery Þlter (see ÒThe DeÞbrillation Recovery ÞlterÓ on pg. 10-6).<br />

A broken wire in a patient lead or a poorly applied sensor may also cause an<br />

ÒOVERLOADÓ message. This condition must be corrected before the overload<br />

condition clears.<br />

Rapid, large and erratic deflections<br />

A broken wire in the patient lead or a poorly applied sensor may cause rapid,<br />

large and erratic trace deßections.<br />

Irregular frequency or amplitude<br />

Patient movement and muscle tremor may result in abnormal traces. To<br />

minimize this artifact, toggle the artifact Þlter to 40 Hz with the Ò9Ó machine<br />

control key.<br />

10-4 Eclipse <strong>Operating</strong> <strong>Instructions</strong>