Chapter 10<br />
Troubleshooting<br />
PROBLEM<br />
Problems using full format paper.<br />
REDUCED PERFORMANCE MODE<br />
Paper speed erratic. ECG printouts have<br />
unexpected breaks in the waveforms at<br />
random intervals. Breaks are<br />
accompanied by vertical dashed lines,<br />
header/footer information and lead<br />
designators.<br />
Report is printing ÒSTATÓ in the Patient<br />
ID Þeld of the report.<br />
PROBABLE CAUSE<br />
- Load paper (see ÒLoading Recording<br />
PaperÓ on pg. 2-6)<br />
- Ensure the paper is at top-of-form by<br />
pressing the ÒPÓ key and waiting until<br />
the paper stops. If ÒPaper OUT:<br />
Reload and continue?Ó message<br />
appears, select OK.<br />
- Press ÒMANÓ. When printing starts,<br />
press ÒSTOPÓ.<br />
- Again, press ÒMANÓ. When printing<br />
starts, press ÒSTOPÓ.<br />
- 50 mm/s paper speed selected.<br />
Reduce paper speed to 25 mm/s.<br />
- Artifact Þlter is set to 40 Hz (on).<br />
Change setting to 150 Hz.<br />
- Current Custom Lead setting is Frank.<br />
Use another Custom Lead setting or<br />
reduce paper speed and turn off<br />
Artifact Filter.<br />
- No ID number was entered for the<br />
patient.<br />
Recognizing and Reducing ECG Artifacts<br />
“OVERLOAD” message<br />
DeÞbrillating the patient will interfere with the Eclipse sensors. This may<br />
result in loss of the trace or erratic trace deßections. An ÒOVERLOADÓ<br />
message alerts you that waveforms may not be accurate. To speed recovery<br />
time after deÞbrillation press the ÒBÓ key to activate the DeÞbrillation<br />
Recovery Þlter (see ÒThe DeÞbrillation Recovery ÞlterÓ on pg. 10-6).<br />
A broken wire in a patient lead or a poorly applied sensor may also cause an<br />
ÒOVERLOADÓ message. This condition must be corrected before the overload<br />
condition clears.<br />
Rapid, large and erratic deflections<br />
A broken wire in the patient lead or a poorly applied sensor may cause rapid,<br />
large and erratic trace deßections.<br />
Irregular frequency or amplitude<br />
Patient movement and muscle tremor may result in abnormal traces. To<br />
minimize this artifact, toggle the artifact Þlter to 40 Hz with the Ò9Ó machine<br />
control key.<br />
10-4 Eclipse <strong>Operating</strong> <strong>Instructions</strong>