Chapter 10<br />
Troubleshooting<br />
PROBLEM<br />
Fax function is not working.<br />
Several functions in the DIRECTORY<br />
menu are unavailable and appear gray.<br />
PROBABLE CAUSE<br />
- Verify that the phone number and<br />
description Þelds are correctly Þlled<br />
in. These are accessed through the<br />
FAX ECG menu.<br />
- Verify that the correct connection type<br />
is selected in the SYSTEM SETUP<br />
menu.<br />
- If MODEM is selected for the<br />
Connection Þeld, verify that the<br />
Phone # Þeld is correctly Þlled in.<br />
These are found in the SYSTEM<br />
SETUP menu.<br />
Troubleshooting Trace Problems<br />
PROBLEM<br />
Waveforms are ßat for all leads and/or<br />
ÒOVERLOADÓ message will not clear.<br />
Waveform is ßat for one or more leads<br />
(all others are OK) and/or<br />
ÒOVERLOADÓ message will not clear.<br />
Baseline is drifting in waveform for one<br />
or more leads.<br />
PROBABLE CAUSE<br />
- Electrodes not applied correctly.<br />
- Acquiring ECG too quickly; wait for<br />
the message, ÒSENSORS OKÓ before<br />
pressing the ECG key.<br />
- Patient cable not properly connected<br />
to unit.<br />
- Faulty patient cable.<br />
- Electrical interference from an<br />
external source is causing noise<br />
spikes.<br />
- Electrode not applied correctly.<br />
- Acquiring ECG too quickly; wait for<br />
the message, ÒSENSORS OKÓ before<br />
pressing the ECG key.<br />
- Electrode not adhering properly.<br />
- Electrode is being pulled, tapped or<br />
pressed.<br />
- Electrode disconnected from patientÕs<br />
skin.<br />
- Lead wire disconnected from<br />
electrode.<br />
- Debris in teeth of astroclip.<br />
- Faulty lead wire(s) or patient cable.<br />
- Poor patient preparation.<br />
- Use of dissimilar sensors or sensors<br />
not recommended for use with<br />
Eclipse .<br />
- Sensors need to sit longer on skin.<br />
- Poor sensor contact with skin.<br />
10-2 Eclipse <strong>Operating</strong> <strong>Instructions</strong>