Troubleshooting<br />

12 Troubleshooting<br />

Symptom Cause Remedy<br />

Periodic superimposition of AC<br />

line interference (60 Hz) (Figure<br />

12-1)<br />

Superimposition of irregular AC<br />

line interference (Figure 12-2)<br />

The printed date and time are<br />

incorrect<br />

The green standby indicator 23<br />

does not light up, although the<br />

recorder is connected to the<br />

power line<br />

The recorder does not write over<br />

the entire paper width<br />

In 12 Lead Mode, the recorder<br />

does not stop and continues to<br />

feed paper. This does not happen<br />

in 6 Lead Mode.<br />

Recorder does not start after<br />

activation of the key, or<br />

the recording is aborted.<br />

interference from the power line<br />

Muscle artifact caused by<br />

patient movements, hiccup,<br />

coughing<br />

Built-in lithium battery is<br />

depleted. The battery has a life<br />

of approx. 5 years<br />

Defective AC power adapter or<br />

fuse<br />

Paper compartment not properly<br />

closed<br />

The paper pad was inserted the<br />

wrong way round so the recorder<br />

cannot detect the cue<br />

mark<br />

Unit operated on battery power:<br />

battery discharged<br />

Ground bed, verify position of the<br />

leadwires, switch on AC filter<br />

The patient should be warm enough<br />

and resting comfortably (place cushions<br />

under arms and knees).<br />

Comfort patient or distract patient's<br />

attention, enable muscle filter (20 Hz /<br />

40 Hz), if necessary.<br />

Notify service to check and/or replace<br />

battery<br />

Notify service to check and/or replace<br />

fuse<br />

Paper door must lock into place on<br />

both sides<br />

Insert the paper pad as instructed<br />

Connect recorder to the power line.<br />

After a few minutes, the recorder is<br />

able to resume operation. Always<br />

connect recorder to the power line<br />

when the battery low indicator 22<br />

lights up. The battery capacity depends<br />

on age, temperature and charge level<br />

(chapter 3 "Putting the Device into<br />

Operation").<br />

No recording in 12 Lead Mode Failure of at least one electrode Check all electrodes or enable Override<br />

function (section 9.2 "12 Lead Mode").<br />

Paper jam<br />

Open paper compartment and removed<br />

jammed sheet,<br />

place beginning of paper between the<br />

marks, close paper compartment and<br />

press .<br />

227 492 04-F <strong>MAC</strong>® <strong>1200</strong> 61

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!