System Setup<br />
9 System Setup<br />
9.1 Some Basic Facts<br />
• Press setup<br />
to display the setup menu.<br />
The setup menu with the following options will<br />
appear:<br />
− Operating mode: 12 Lead (6 Lead, Arrhythmia)<br />
− General Settings<br />
− Communication<br />
− Patient Data Setup<br />
− Option Code<br />
At "Operating mode", you will always see the currently<br />
selected mode. So be sure to select the appropriate<br />
mode before entering the setup menu.<br />
• To access the menu options, position the bar<br />
cursor on the option with the cursor keys and<br />
confirm the selection with .<br />
The operating steps to select a setting are always the<br />
same:<br />
• Using the cursor keys and , you select the<br />
option, then you confirm the selection with .<br />
The cursor will move to the next menu item.<br />
• Individual items can be skipped with or .<br />
• Press to exit the setup mode.<br />
setup<br />
format/<br />
speed<br />
J % K ( L ) muscle<br />
X<br />
filter<br />
lead<br />
6<br />
N < M > store/<br />
gain retrieve arrhy<br />
U 7 I 8 O 9 P 0 start<br />
pat<br />
info<br />
copy<br />
12<br />
stop<br />
9.2 12 Lead Mode<br />
• Use the cursor keys to position the bar cursor on<br />
"12 Lead" and confirm the selection with .<br />
The setup menu for automatic 12 lead recording will<br />
appear.<br />
The angular brackets [ ] identify the system<br />
defaults.<br />
Report Sequence<br />
[STANDARD] (I, II, III, aVR, aVL, aVF, V1, V2,<br />
V3, V4, V5, V6)<br />
CABRERA (aVL, I, -aVR, II, aVF, III, V1, V2, V3,<br />
V4, V5, V6)<br />
Rhythm Leads<br />
Any three of the available ECG leads can be selected<br />
as rhythm leads. They are printed with report formats<br />
4x2.5R3 and 1x10R3. Formats 2x5R1 and<br />
4x2.5R1 will show only the first of the rhythm leads.<br />
Gain<br />
5, [10], 20, 40 mm/mV, *auto<br />
Report Format<br />
For an overview of the available report formats,<br />
refer to section 5.4 "The Report Formats". The<br />
default format is 4x2.5R1.<br />
Detailed Results<br />
The "Detailed results" page will be printed,<br />
Yes/[No] (section 5.4 "The Report Formats", only<br />
available with option MEAS or DIAG).<br />
standby battery low<br />
24 8 9<br />
Figure 9-1. System setup keys<br />
19<br />
Muscle Filter/AC Filter<br />
Elimination of muscle artifact and AC line interference.<br />
Default: muscle filter [No], AC filter [Yes]<br />
48 <strong>MAC</strong>® <strong>1200</strong> 227 492 04-F