MAC ® 1200 Operations Manual - Jaken Medical...
MAC ® 1200 Operations Manual - Jaken Medical...
MAC ® 1200 Operations Manual - Jaken Medical...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Arrhythmia Mode<br />
7.2 Recording<br />
*RL*: right leg electrode disconnected<br />
*RA*: right arm electrode disconnected<br />
*LA*: left arm electrode disconnected<br />
*LL*: left leg electrode disconnected<br />
*V1*: chest electrode V1 disconnected<br />
*V2*: chest electrode V2 disconnected<br />
*V3*: chest electrode V3 disconnected<br />
*V4*: chest electrode V4 disconnected<br />
*V5*: chest electrode V5 disconnected<br />
*V6*: chest electrode V6 disconnected<br />
Messages indicating disconnected electrodes<br />
Note<br />
- With copy , a single-page recording can be initiated<br />
after program start.<br />
- Please note that filters may suppress diagnostically<br />
relevant portions of the signal, because they<br />
limit the transmission range.<br />
Filters should therefore only be enabled if necessary.<br />
- For proper functioning of the ECG analysis algorithm,<br />
pacemaker patients must be identified<br />
in the patient data: Pacemaker – Yes (section 4.4<br />
"Entering Patient Data").<br />
− After switching on the unit, press<br />
the Arrhythmia Mode.<br />
arrhy<br />
to select<br />
− Before recording the ECG, patient data can be<br />
pat<br />
entered (<br />
info<br />
). We recommend to enter the patient's<br />
name to annotate it on each page.<br />
− Before initiating a recording, check the display<br />
for error messages (see table at left). Check all<br />
electrodes; if the message persists, there must be<br />
a break in the patient cable. Replace the cable<br />
with a new one.<br />
− The recording is started and stopped with the<br />
key.<br />
Upon program start, the unit records 6 leads of ECG<br />
(1 page). During the following learn phase, the<br />
analysis algorithm learns the patient's typical QRS<br />
complex. After the learn phase, the recorder prints a<br />
report where the QRS complexes acquired in the<br />
learn phase are labeled "L" and the complex found<br />
to be the patient's typical complex is labeled<br />
"QRSL". Having completed the learn phase, the<br />
<strong>MAC</strong> <strong>1200</strong> is ready to identify arrhythmias.<br />
With the system defaults, the <strong>MAC</strong> <strong>1200</strong> will activate<br />
the following functions and settings:<br />
− the STD_C report sequence V1 through V6<br />
(configurable)<br />
− a gain of 10 mm/mV (configurable) (calibration<br />
pulse at the beginning of the recording). The unit<br />
can be set up to automatically adapt the gain to<br />
the ECG signal (*auto)<br />
− the AC filter is enabled (configurable)<br />
− the muscle filter is disabled (configurable)<br />
44 <strong>MAC</strong>® <strong>1200</strong> 227 492 04-F