Arrhythmia Mode<br />
7 Arrhythmia Mode<br />
GE marquette <strong>MAC</strong> <strong>1200</strong> John Doe<br />
7.1 Some Basic Facts<br />
In Arrhythmia Mode, the <strong>MAC</strong> <strong>1200</strong> continuously<br />
scans the ECG for arrhythmias.<br />
Figure 7-1. Event recording<br />
Note<br />
After starting the program, press format/<br />
speed to select a<br />
continuous recording with a speed of 5 mm/s (configurable).<br />
If the unit identifies an arrhythmic event,<br />
it will automatically switch to the fast paper speed.<br />
With the same key format/<br />
speed<br />
, the trend recording can be<br />
stopped. The unit can be set up to automatically start<br />
a trend recording when the Arrhythmia Mode is<br />
initiated.<br />
Parameter<br />
Report sequence<br />
System defaults Options<br />
STD_C (chest<br />
leads V1<br />
through V6)<br />
STD_RED (I, II,<br />
III, V2, V4, V6)<br />
STD_LI, (I, II, III,<br />
aVR, aVL, aVF)<br />
CABR_LI (aVL, I,<br />
-aVR, II, aVF, III)<br />
HIGH_C (V1'<br />
through V6')<br />
Gain 10 mm/mV "*auto", 5, 20, 40<br />
mm/mV<br />
Muscle Filter No Yes<br />
Muscle Filt. Freq. 40 Hz 20 Hz<br />
AC Filter Yes No<br />
Trend Rec. No Yes<br />
Arrhy Data unequal all, no<br />
Episodes chron. prio, ventr., no<br />
From six simultaneously acquired leads, the <strong>MAC</strong><br />
<strong>1200</strong> automatically selects the two that provide the<br />
best signal for analysis.<br />
When the analysis algorithm detects an arrhythmia,<br />
the event is recorded with "context" (Figure 7-1).<br />
The length of the recording varies with the duration<br />
of the event episode. In the setup menu (section 9.4<br />
"Arrhythmia Mode") you determine the conditions<br />
for a recording:<br />
− the recorder starts each time it detects a singlebeat<br />
event - all<br />
− the recorder starts each time it detects an event<br />
different from the previous event - unequal<br />
− the recorder does not start at all - no.<br />
Some of the system settings can be customized.<br />
They are labeled with "configurable". The following<br />
information refers to a unit with the system defaults<br />
(see table at left). For instructions on changing the<br />
system setup, refer to section 9.4 "Arrhythmia<br />
Mode".<br />
227 492 04-F <strong>MAC</strong>® <strong>1200</strong> 43