Recording in 6 Lead Mode<br />
− With / you toggle between the two lead<br />
sets (3 each) on the display that belong to the recorded<br />
group.<br />
− When the anti-drift system is enabled, there will<br />
be a short delay before the recording starts. The<br />
ECG will then be recorded with a delay of 2.2 s.<br />
The heart rate limit is automatically calculated from<br />
the date of birth (WHO 100% = 220 - age). When<br />
the date of birth is not entered, the unit will set the<br />
limit at 180 bpm. This value can be changed with<br />
and ( in steps of 5 bpm). The minimum<br />
value for the heart rate limit is 30 bpm.<br />
6.3 Brief Operating Instructions –<br />
6 Lead Mode<br />
• Switch on the unit and wait for self-test to end<br />
• Apply electrodes to patient<br />
• Select the 6 Lead Mode -<br />
• Enter patient data -<br />
• Check device settings<br />
- report sequence<br />
- AC filter<br />
- ADS (cubic spline)<br />
pat<br />
info<br />
- heart rate alarm limits<br />
6<br />
• Modify device settings, if required - setup<br />
• Watch ECG on display<br />
• Check that no lead failure message is displayed<br />
• Start recording with<br />
• Proceed to the next group of 6 leads with lead<br />
• Change the writer speed with format/<br />
speed<br />
• Switch on muscle filter with muscle<br />
filter<br />
• Stop the recording with<br />
• Print patient data with copy<br />
42 <strong>MAC</strong>® <strong>1200</strong> 227 492 04-F