Recording in 6 Lead Mode<br />
6 Recording in 6 Lead Mode<br />
6.1 Some Basic Facts<br />
In 6 Lead Mode, the system acquires 6 leads of ECG<br />
in realtime. Recordings are started and stopped with<br />
. Some of the system settings can be customized.<br />
They are labeled with "configurable".<br />
The following information refers to a unit with the<br />
system defaults (see table below). For instructions<br />
on changing the device setup, refer to section 9.3<br />
"6 Lead Mode".<br />
Parameter System defaults Options<br />
Report Sequence STANDARD CABRERA,<br />
SEQ.NO.4<br />
Gain 10 mm/mV "*auto", 5, 20, 40<br />
mm/mV<br />
Speed 25 mm/s 5, 50 mm/s<br />
Muscle Filter No Yes<br />
Muscle Filt.<br />
Frequ.<br />
40 Hz 20 Hz<br />
AC Filter Yes No<br />
Anti-Drift System No Yes<br />
Auto. Paper Feed Yes No<br />
6.2 Recording<br />
After switching on the unit, press<br />
6 Lead Mode.<br />
6<br />
to select the<br />
- Before recording the ECG, patient data can be<br />
pat<br />
entered with<br />
info<br />
. We recommend to enter the<br />
patient's name to annotate it on each page.<br />
Note<br />
In 6 Lead Mode, messages indicating disconnected<br />
electrodes are annotated on the recording, e.g. Lead<br />
fail V1.<br />
− Before initiating a recording, check the display<br />
for error messages (see table below). Check all<br />
electrodes; if the message persists, there must be<br />
a break in the patient cable. Replace the cable<br />
with a new one.<br />
*RL*: right leg electrode disconnected<br />
*RA*: right arm electrode disconnected<br />
*LA*: left arm electrode disconnected<br />
*LL*: left leg electrode disconnected<br />
*V1*: chest electrode V1 disconnected<br />
*V2*: chest electrode V2 disconnected<br />
*V3*: chest electrode V3 disconnected<br />
*V4*: chest electrode V4 disconnected<br />
*V5*: chest electrode V5 disconnected<br />
*V6*: chest electrode V6 disconnected<br />
Messages indicating disconnected electrodes<br />
40 <strong>MAC</strong>® <strong>1200</strong> 227 492 04-F