Recording in 6 Lead Mode<br />

6 Recording in 6 Lead Mode<br />

6.1 Some Basic Facts<br />

In 6 Lead Mode, the system acquires 6 leads of ECG<br />

in realtime. Recordings are started and stopped with<br />

. Some of the system settings can be customized.<br />

They are labeled with "configurable".<br />

The following information refers to a unit with the<br />

system defaults (see table below). For instructions<br />

on changing the device setup, refer to section 9.3<br />

"6 Lead Mode".<br />

Parameter System defaults Options<br />

Report Sequence STANDARD CABRERA,<br />

SEQ.NO.4<br />

Gain 10 mm/mV "*auto", 5, 20, 40<br />

mm/mV<br />

Speed 25 mm/s 5, 50 mm/s<br />

Muscle Filter No Yes<br />

Muscle Filt.<br />

Frequ.<br />

40 Hz 20 Hz<br />

AC Filter Yes No<br />

Anti-Drift System No Yes<br />

Auto. Paper Feed Yes No<br />

6.2 Recording<br />

After switching on the unit, press<br />

6 Lead Mode.<br />

6<br />

to select the<br />

- Before recording the ECG, patient data can be<br />

pat<br />

entered with<br />

info<br />

. We recommend to enter the<br />

patient's name to annotate it on each page.<br />

Note<br />

In 6 Lead Mode, messages indicating disconnected<br />

electrodes are annotated on the recording, e.g. Lead<br />

fail V1.<br />

− Before initiating a recording, check the display<br />

for error messages (see table below). Check all<br />

electrodes; if the message persists, there must be<br />

a break in the patient cable. Replace the cable<br />

with a new one.<br />

*RL*: right leg electrode disconnected<br />

*RA*: right arm electrode disconnected<br />

*LA*: left arm electrode disconnected<br />

*LL*: left leg electrode disconnected<br />

*V1*: chest electrode V1 disconnected<br />

*V2*: chest electrode V2 disconnected<br />

*V3*: chest electrode V3 disconnected<br />

*V4*: chest electrode V4 disconnected<br />

*V5*: chest electrode V5 disconnected<br />

*V6*: chest electrode V6 disconnected<br />

Messages indicating disconnected electrodes<br />

40 <strong>MAC</strong>® <strong>1200</strong> 227 492 04-F