Recording in 12 Lead Mode<br />

5.5.5 Receiving Data with the CSI Communication<br />

Protocol<br />

Initializing modem . . . .<br />

Cancel : [ START / STOP]<br />

Figure 5-15. Screen display after activation of<br />

and copy<br />

Modem initialized<br />

Waiting for ECG data : [ ENTER ]<br />

Cancel : [ START / STOP]<br />

Figure 5-16. Screen display after modem initialization<br />

12Lead/REC 60Hz ADS<br />

4x2.5R1 10mm/mV Standard<br />

Figure 5-17. Unit is ready to receive data<br />

Receiving ECG . . . . 1<br />

Cancel : [START / STOP]<br />

(see also chapter 13 "Technical Specifications")<br />

Receiving ECGs is only possible in the 12Lead<br />

Mode.<br />

• Use the key combination and copy to display<br />

the screen for receiving ECGs (Figure 5-15). The<br />

connected modem is automatically initialized.<br />

The procedure can be aborted with .<br />

• Press to enable the "receive data" mode. The<br />

procedure can be aborted with .<br />

When you have enabled the "receive data" mode, the<br />

standard screen display of the 12 Lead Mode displays.<br />

The message "12Lead/REC" indicates that the<br />

unit is ready to receive data (Figure 5-17).<br />

ECGs can be recorded in the 12 Lead Mode even<br />

while the unit is in the "receive data" mode.<br />

A message displays on the screen when the unit is<br />

receiving data (Figure 5-18). The reception can be<br />

aborted with .<br />

The ECG which has just been received is processed<br />

for the printout. The report is printed in the selected<br />

format. Multiple ECGs are received and printed one<br />

after the other.<br />

After printout of the last ECG, the "receive data"<br />

mode is automatically disabled. The mode is also<br />

disabled when you select another operating mode.<br />

The following information is annotated in the bottom<br />

line of each report:<br />

– the sender<br />

– the software version and analysis program version<br />

used at the sending unit (e.g. ACQ-DEV:<br />

M<strong>1200</strong> V5.1M12i 12SL V13).<br />

Figure 5-18. Unit is receiving ECG 1<br />

227 492 04-F <strong>MAC</strong>® <strong>1200</strong> 37