MAC ® 1200 Operations Manual - Jaken Medical...
MAC ® 1200 Operations Manual - Jaken Medical...
MAC ® 1200 Operations Manual - Jaken Medical...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Recording in 12 Lead Mode<br />
Modem Error Messages<br />
ECG Transmission Error! (CSI)<br />
Check interface!<br />
No dial tone!<br />
Busy!<br />
No answer!<br />
No carrier!<br />
Check modem configuration!<br />
Cause<br />
The connection was interrupted due to a fault.<br />
Fault in RS232 interface or modem. Modem may be<br />
switched off.<br />
No dial tone detected.<br />
Busy signal detected.<br />
No answer at remote end.<br />
Carrier signal lost or not detected.<br />
Modem configuration error.<br />
34 <strong>MAC</strong>® <strong>1200</strong> 227 492 04-F