Recording in 12 Lead Mode<br />

5.4 The Report Formats<br />

The length and scope of the reports depends on the<br />

implemented software (standard, MEAS (measurement),<br />

DIAG (interpretation)).<br />

The table below shows all of the 6 different report<br />

formats available with the <strong>MAC</strong> <strong>1200</strong> units.<br />

Format ECG traces Rhythm lead Speed Measurement* Interpretation* Pages<br />

length/leads length/leads<br />

4x2.5R1 (default 4x2.5s/4x3 10 s/1 25 mm/s Yes Yes 1<br />

format)<br />

4x2.5R3 4x2.5 s/4x3 10 s/3 25 mm/s Yes Yes 1<br />

2x5R1 2x5 s/2x6 10 s/1 25 mm/s Yes Yes 1<br />

2x5_50 2x5 s/2x6 No 50 mm/s Yes Yes 2<br />

1x10R12 10 s/1x12 No 25 mm/s No No 1<br />

1x10R3 10 s/1x3 10 s/3 25 mm/s Yes Yes 1<br />

* measurement results and interpretative statements are only available from <strong>MAC</strong> <strong>1200</strong> with the appropriate<br />

software options<br />

Note<br />

- The printed reports are unconfirmed documents.<br />

They must be overread, verified, and signed by a<br />

physician for confirmation.<br />

- The heart rate HR annotated on the report pages<br />

is calculated from all beats of the 10 second<br />

ECG.<br />

- To obtain a printout of the full patient data, select<br />

the 6 Lead Mode and press copy .<br />

Detailed results<br />

In the setup menu of units equipped with the MEAS<br />

or DIAG option, you can choose the "Detailed<br />

results" page. When selected, this page will be<br />

appended to the reports. It contains patient data,<br />

measurement results (MEAS), interpretative statements<br />

(DIAG), medians and the tabular measurement<br />

values.<br />

GEMS IT <strong>MAC</strong> <strong>1200</strong> John Doe<br />

University Hospital, Dr. Williams<br />

male, Caucasian, 32 years, 6.7 ft., 172 lb, 161/133 mmHg Pacemaker 414 355 378<br />

HR<br />

101BPM<br />

Feb.09.2001 11:07:22 AM 25mm/s 10mm/mV ADS 50Hz 0.08 - 150Hz 1x10R12 12 Lead 12SL V5.2<br />

Figure 5-5. 1x10R12 report format<br />

227 492 04-F <strong>MAC</strong>® <strong>1200</strong> 31