Recording in 12 Lead Mode<br />

5 Recording in 12 Lead Mode<br />

Note<br />

Please bear in mind that no automated analysis of<br />

ECG signals is completely reliable. Therefore a physician<br />

should always overread and reassess the system<br />

interpretation before performing patient diagnosis.<br />

5.1 Some Basic Facts<br />

In 12 Lead Mode, 12 leads of ECG are acquired<br />

simultaneously for a period of 10 seconds. When<br />

initiated with , ECG acquisition and recording<br />

proceed automatically. The system, however,<br />

may be set up to start recording only when<br />

specific patient data (ID, Secondary ID, name) have<br />

been entered (see section 9.7 "Configuration of<br />

Patient Data Menu").<br />

Depending on the implemented software options, the<br />

ECG<br />

− is only printed out (options MEAS - measurement<br />

-, DIAG - interpretation - not implemented)<br />

− is measured and printed out with the measurement<br />

results (with option MEAS - measurement)<br />

− is measured, interpreted (analyzed) and printed<br />

out with the interpretative statements (with option<br />

DIAG - interpretation)<br />

Units equipped with the optional Memory function<br />

MEMO can save up to 40 resting ECG. These ECGs<br />

can be<br />

− printed or<br />

− sent to the MUSE CV Information System (CSI<br />

protocol) (see section 5.3 "The Storage Program").<br />

The unit offers different report formats for printout<br />

of the ECG. With the system defaults, all 12 leads<br />

including the measurement and analysis results will<br />

be documented on a single page (see section 5.4<br />

"The Report Formats").<br />

Several system settings can be customized. In this<br />

manual they are labeled "configurable".<br />

The following information refers to a unit with the<br />

system defaults (see table below). For instructions<br />

on changing the system setup, refer to section 9.2<br />

"12 Lead Mode".<br />

Parameter<br />

System defaults<br />

Options<br />

Report Sequence STANDARD CABRERA<br />

Rhythm Leads II, V1, V5 I, III,<br />

aVR, aVL, aVF,<br />

V2, V3, V4, V6<br />

Gain 10 mm/mV "*auto", 5, 20, 40<br />

mm/mV<br />

Report Format 4x2.5R1 1x10R12, 2x5R1,<br />

2x5_50, 4x2.5R1,<br />

1x10R3, 4x2.5R3<br />

Detailed Results No Yes<br />

Muscle Filter No Yes<br />

Muscle Filt. Freq. 40 Hz 20 Hz<br />

AC Filter Yes No<br />

<strong>Manual</strong> copy to EKG HOST<br />

No. of copies 1 0 to 9<br />

Delete ECG after<br />

Transm.<br />

No<br />

Yes<br />

Autosave ECG No Yes<br />

Use Screening<br />

criteria<br />

Suppress<br />

NORMAL statem.<br />

Suppress<br />

ABNORMAL<br />

statm.<br />

No<br />

No<br />

No<br />

Yes<br />

Yes<br />

Yes<br />

Interpretation Yes No<br />

Print Interpretation<br />

Yes<br />

Override Function Yes<br />

No<br />

No<br />

26 <strong>MAC</strong>® <strong>1200</strong> 227 492 04-F