Recording in 12 Lead Mode<br />
5 Recording in 12 Lead Mode<br />
Note<br />
Please bear in mind that no automated analysis of<br />
ECG signals is completely reliable. Therefore a physician<br />
should always overread and reassess the system<br />
interpretation before performing patient diagnosis.<br />
5.1 Some Basic Facts<br />
In 12 Lead Mode, 12 leads of ECG are acquired<br />
simultaneously for a period of 10 seconds. When<br />
initiated with , ECG acquisition and recording<br />
proceed automatically. The system, however,<br />
may be set up to start recording only when<br />
specific patient data (ID, Secondary ID, name) have<br />
been entered (see section 9.7 "Configuration of<br />
Patient Data Menu").<br />
Depending on the implemented software options, the<br />
ECG<br />
− is only printed out (options MEAS - measurement<br />
-, DIAG - interpretation - not implemented)<br />
− is measured and printed out with the measurement<br />
results (with option MEAS - measurement)<br />
− is measured, interpreted (analyzed) and printed<br />
out with the interpretative statements (with option<br />
DIAG - interpretation)<br />
Units equipped with the optional Memory function<br />
MEMO can save up to 40 resting ECG. These ECGs<br />
can be<br />
− printed or<br />
− sent to the MUSE CV Information System (CSI<br />
protocol) (see section 5.3 "The Storage Program").<br />
The unit offers different report formats for printout<br />
of the ECG. With the system defaults, all 12 leads<br />
including the measurement and analysis results will<br />
be documented on a single page (see section 5.4<br />
"The Report Formats").<br />
Several system settings can be customized. In this<br />
manual they are labeled "configurable".<br />
The following information refers to a unit with the<br />
system defaults (see table below). For instructions<br />
on changing the system setup, refer to section 9.2<br />
"12 Lead Mode".<br />
Parameter<br />
System defaults<br />
Options<br />
Report Sequence STANDARD CABRERA<br />
Rhythm Leads II, V1, V5 I, III,<br />
aVR, aVL, aVF,<br />
V2, V3, V4, V6<br />
Gain 10 mm/mV "*auto", 5, 20, 40<br />
mm/mV<br />
Report Format 4x2.5R1 1x10R12, 2x5R1,<br />
2x5_50, 4x2.5R1,<br />
1x10R3, 4x2.5R3<br />
Detailed Results No Yes<br />
Muscle Filter No Yes<br />
Muscle Filt. Freq. 40 Hz 20 Hz<br />
AC Filter Yes No<br />
<strong>Manual</strong> copy to EKG HOST<br />
No. of copies 1 0 to 9<br />
Delete ECG after<br />
Transm.<br />
No<br />
Yes<br />
Autosave ECG No Yes<br />
Use Screening<br />
criteria<br />
Suppress<br />
NORMAL statem.<br />
Suppress<br />
ABNORMAL<br />
statm.<br />
No<br />
No<br />
No<br />
Yes<br />
Yes<br />
Yes<br />
Interpretation Yes No<br />
Print Interpretation<br />
Yes<br />
Override Function Yes<br />
No<br />
No<br />
26 <strong>MAC</strong>® <strong>1200</strong> 227 492 04-F