Preparations for ECG Recording<br />
4.3 Artifact Due to Poor Electrode Application<br />
The resting ECG analysis system is equipped with<br />
state-of-the-art electronic utilities that ensure artifact-free<br />
recordings. Among these are the automatic<br />
baseline adjustment and the anti-drift system (ADS<br />
or cubic spline).<br />
At the beginning of the recording the automatic<br />
baseline adjustment algorithm verifies the incoming<br />
signal and adjusts the baseline position accordingly.<br />
During the recording, the anti-drift system (cubic<br />
spline) continuously checks the baseline position<br />
and returns it to the normal level, if required (Figure<br />
4-6).<br />
For the 6 Lead Mode, the anti-drift system (cubic<br />
spline) can be enabled and disabled from the setup<br />
menu, in the 12 Lead and Arrhythmia Modes, it is<br />
always enabled.<br />
When electrodes are not properly applied, these<br />
measures may not fully compensate for artifact.<br />
High polarization voltages induced by electrodes<br />
applied without conductive gel may cause the amplifier<br />
to overrange, so that a straight line will be<br />
recorded instead of the ECG (see Figure 4-6). The<br />
device will then automatically block and return the<br />
baseline to its normal position. A baseline is then<br />
recorded for approx. 1 second. It is possible to block<br />
the amplifiers manually by disconnecting the RL<br />
electrode.<br />
On the display this condition is indicated by ****<br />
instead of the electrode label (e.g. at i, Figure 5-1).<br />
Remedy<br />
• Apply the electrodes according to instructions.<br />
• Do not apply the electrodes on top of clothing.<br />
• Use a contact agent with reusable electrodes (e.g.<br />
moistened electrode paper, electrode cream,<br />
spray, etc.).<br />
• Wait approx. 10 seconds before initiating a<br />
recording. After the 10-second period, the automatic<br />
functions are enabled and the polarization<br />
voltages have stabilized, provided the electrodes<br />
are properly applied. In case of improper electrode<br />
application, an error message will appear<br />
on the display (RL, LL, LA, LL, V1 to V6).<br />
• If required, the ADS (cubic spline) and the filters<br />
(20/40 Hz, 60 Hz) can be disabled to verify the<br />
“raw” ECG signal.<br />
approx.1 s<br />
Figure 4-6. Sample recording<br />