Putting the Device into Operation and Performance Test<br />

Parameter<br />

Ordering Physician<br />

Referring Physician<br />

System Defaults Options<br />

No<br />

No<br />

selection from a list<br />

of 10 names<br />

selection from a list<br />

of 10 names<br />

Technician No selection from a list<br />

of 10 names<br />

Institution Name empty text box text box (40 chrs)<br />

Cart # 1 1 to 9999<br />

Site #(for CSI<br />

protocol only)<br />

1 1 to 255<br />

Location # blank 1 to 600<br />

Date<br />

(dd.mm.yyyy)<br />

Time (hh:mm)<br />

current date<br />

current time<br />

Lead Fail Beep No Yes<br />

High HR Beep No Yes<br />

Lead Labels AAMI IEC<br />

Pace Enhancement<br />

Yes No<br />

Baseline roll filter 0.08 Hz 0.04 Hz, 0.16 Hz<br />

Date mm/dd/yyyy dd.mm.yyyy<br />

Time 12 24<br />

Units in, lb cm, kg<br />

Mains 60 Hz 50 Hz<br />

LCD light off<br />

after<br />

5 min 1 to 99 minutes<br />

Low battery beep 0 (Off) 5 s to 60 s (5 s =<br />

beep at 5-s interval)<br />

Default mode 12 Lead 6 Lead,<br />

Arrhythmia<br />

Language English English, French,<br />

Spanish<br />

Enable password No Yes<br />

Test DATA No Yes<br />

Restore defaults No Yes<br />

Print setup lists No Yes<br />

3.5 System Setup<br />

The table at left shows the system setup parameters<br />

that can be modified and the system defaults.<br />

For instructions on changing the device setup, refer<br />

to section 9.5 "System Setup".<br />