Putting the Device into Operation and Performance Test<br />

3.3 Installation and Mains Connection<br />

Figure 3-2 shows a practical arrangement of patient<br />

and recorder. For interference-free operation, it is<br />

important that the patient cable and the power cord<br />

do not run parallel.<br />

Figure 3-2. Arranging device and couch<br />

• Using the power cord, connect the device to the<br />

power line (Figure –3-1). Use only the original<br />

power cord or an equivalent cable.<br />

The standby indicator 23 will illuminate.<br />

• Check the paper supply (the window in the paper<br />

door allows you to look inside the compartment).<br />

If it is necessary to insert a new paper pad, refer<br />

to chapter 10 for instructions.<br />

3.4 Performance Check<br />

Figure 3-3. AC power input<br />

on<br />

stdby<br />

Q 1 W 2 E 3 R 4 T 5 Y 6<br />

A !<br />

S ?<br />

D =<br />

F -<br />

G +<br />

H *<br />

Z ; X : C / V , B . N <<br />

alt<br />

Figure 3-4. Power switch<br />

Note<br />

– When turning off the <strong>MAC</strong> <strong>1200</strong> (standby mode),<br />

be sure to press the power switch long enough.<br />

– The backlighting of the display switches off<br />

automatically when no key is activated for 5<br />

minutes (adjustable).<br />

– Run the full self-test at least once a day to ensure<br />

that the device is functioning properly.<br />

• Press the power switch to switch on the device<br />

(Figure 3-4).<br />

The amber stop indicator<br />

20 will illuminate.<br />

After power-up, the resting ECG analysis system<br />

runs an automatic self-test. When no problem is<br />

detected, it defaults to the 12 Lead Mode. If a malfunction<br />

is identified, the display will show an error<br />

message “Error...”. In this situation, notify service to<br />

check and repair the device.<br />

The self-test can be aborted with the R 4<br />

button. In<br />

this case, the device immediately activates the 12<br />

Lead Mode.<br />

Contrast Adjustment<br />

• To adjust the contrast, simultaneously press<br />

and the appropriate cursor key:<br />

for more contrast, for less contrast.<br />

alt<br />

227 492 04-F <strong>MAC</strong>® <strong>1200</strong> 17