Putting the Device into Operation and Performance Test<br />

3.2 Power Supply<br />

standby<br />

battery low<br />

format/<br />

speed<br />

pat<br />

info<br />

store/<br />

retrieve<br />

The units are powered from the power line or from<br />

the rechargeable battery.<br />

The battery charges automatically when the unit is<br />

connected to the power line. It is not necessary to<br />

turn on the device for charging. As soon as the<br />

device is connected to the power line, the standby<br />

indicator 23 lights up (Figure 3-3). To ensure that<br />

the battery is always fully charged, leave the <strong>MAC</strong><br />

<strong>1200</strong> resting ECG analysis system connected to the<br />

power line whenever possible. After 4 hours the<br />

battery has regained its full capacity.<br />

23<br />

22<br />

Figure 3-1. Indicators<br />

The battery low indicator 22 is illuminated when<br />

battery needs to be charged. The unit can also be set<br />

up to emit an additional audio signal when the<br />

battery requires charging.<br />

With a full battery, about 50 ECGs (1 page) can be<br />

recorded in 12 Lead Mode. When its capacity drops<br />

to about 25 recordings, the battery is used up and<br />

must be replaced by a service specialist.<br />

Note<br />

To prolong the battery life, discharge the battery at<br />

least once per month (by operating the resting ECG<br />

analysis system on battery power).<br />

Note<br />

In standby mode, a fully charged battery is drained<br />

within approx. 4 hours. Therefore, when operating<br />

the device on battery power, be sure to turn it off<br />

when it is not in use.<br />

16 <strong>MAC</strong>® <strong>1200</strong> 227 492 04-F