Resting ECG: Post Test Review<br />

F<br />

G<br />

H<br />

I<br />

Select if you wish to adjust the T-wave offset mark (see “Adjusting the T-offset<br />

Mark” on page 5-19).<br />

Measurement mark.<br />

Menu for selection of the leads to be reanalyzed.<br />

Table showing time intervals and position of the cursor in mV.<br />

Performing Reanalysis<br />

1. Click the measurement mark and hold the mouse button depressed.<br />

2. Move the measurement mark to its new position and release the<br />

mouse button.<br />

3. Adjust the other marks in the same way.<br />

4. Click ECG Analysis.<br />

5. Click Reset and ECG Analysis if you wish to restore the previous<br />

analysis results.<br />

Note<br />

The P-onset and P-offset markers may be dashed, e.g. in the presence<br />

of atrial fibrillation. This indicates that the program was unable to<br />

locate the exact beginning and end of the P wave. In this case you can<br />

click the marks and drag them to the correct position; the marks are<br />

now solid instead of dashed. It is possible to restore the marks to the<br />

original position by clicking them again.<br />

5-18 <strong>CardioSoft</strong> Revision A<br />