Product Overview: Putting the System Into Service<br />

Putting the System Into Service<br />

The software is a Windows-based program. Its scope can be expanded by<br />

activation of individual optional programs. This manual describes the<br />

program with all options included. If you have not purchased one or the<br />

other option, simply skip the corresponding sections.<br />

Access to the program is restricted to registered users. At the time a user<br />

is registered, the password and privileges are assigned (technician,<br />

physician, with or without the right to edit data, etc.).<br />

Mains Connection, Power Up, Functional Test<br />

1. Before powering the system up for the first time, read the safety<br />

information in section “Safety Information” on page 1-6.<br />

2. Turn on the PC, the monitor and, if used, the bicycle ergometer or<br />

treadmill, electrocardiograph and printer.<br />

The program will start up automatically. When you see the initial screen<br />

(see “Initial Screen” on page 2-13) and no error message appears, the<br />

system is operational.<br />

Note<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

Do not run more than one additional Windows-based program at<br />

the same time as <strong>CardioSoft</strong>.<br />

The screen saver is automatically disabled during ECG<br />

acquisition.<br />

When using the program for the first time, select the size of your<br />

monitor screen (“General Tab” on page 11-41).<br />

Disable all energy save modes (BIOS, Windows).<br />

Connect the printer(s) to parallel ports only.<br />

Unlock the purchased optional software programs as described in<br />

section “Option Code Tab” on page 11-57.<br />

Scan the hard drive for viruses once a week.<br />

Do not turn off the PC until you have properly exited from the<br />

program and from Windows.<br />

Run a functional test on a regular basis (about once a month);<br />

this includes an inspection of the cables and other accessories for<br />

signs of damage.<br />

Warning<br />

Simulated Patient Data — To avoid the possibility of<br />

misinterpreting patient information, enable the<br />

simulated Test Patient data for demonstration purposes<br />

only.<br />