Product Overview: PC Setup<br />

Warning<br />

Patient Hazard — Always connect the CAM-14<br />

acquisition module to the CAM-USB interface box, never<br />

directly to the PC.<br />

Always attach the CAM-USB interface box to a suitable<br />

place (e.g. on the PC).<br />

Shock Hazard / Equipment Damage — The patient signal<br />

input is a high-insulation port and it is defibrillationproof<br />

(CAM-14, type BF / CORINA, type CF). This type of<br />

input guarantees patient safety and protects the<br />

equipment during defibrillation and HF surgery.<br />

Nevertheless, extreme care should be exercised when a<br />

defibrillator or HF surgery equipment is used on a<br />

patient connected to other devices. As a general rule, the<br />

distance between the ECG electrodes and the<br />

defibrillation or HF surgery electrodes should not be less<br />

than 15 cm. If this is not ensured, temporarily disconnect<br />

the patient cables while using the defibrillator or the HF<br />

surgery equipment. It is very important that conductive<br />

parts, such as connectors, electrodes, transducers, do not<br />

come into contact with other grounded, conductive parts<br />

when connected to the isolated patient input. Such<br />

contact would bridge the patient's isolation and cancel<br />

the protection provided by the isolated input. In<br />

particular, there must be no contact of the neutral<br />

electrode and ground.<br />

Note<br />

Set up the device so that the operator has a clear,<br />

unobstructed view of the control panel.<br />

Only persons with the required training and expertise are<br />

authorized to use medical electrical equipment such as a PC<br />

with <strong>CardioSoft</strong>.<br />

Revision A <strong>CardioSoft</strong> 2-7<br />

2023324-053