Product Overview: PC Setup<br />

Caution<br />

Proper Leadwire Connection — Improper connection will<br />

cause inaccuracies in the ECG.<br />

Trace each individual leadwire from its acquisition<br />

module label to the colored connector and then to the<br />

proper electrode to ensure that it is matched to the<br />

correct label location.<br />

4. Connect the CAM-USB or the CAM-USB A/T interface box to the<br />

USB port of your PC. Do not connect the CAM-USB interface box to a<br />

USB hub or an extension cable.<br />

5. Connect the AC power adapter to the CAM-USB A/T interface box.<br />

Pull back the coupling sleeve to disconnect.<br />

6. Suction tubing connects here (CAM-USB A/T KISS interface box<br />

only).<br />

7. Trigger and analog signal outputs.<br />

Note<br />

The ECG signal input is isolated and defibrillation-proof if used<br />

in conjunction with the CAM-14 acquisition module / USB or<br />

USB A/T interface box (type BF).<br />

Before starting the program, check that the CAM-USB or CAM-<br />

USB A/T interface box is connected to the PC. If this is not the<br />

case, the system cannot read the internal serial number and all<br />

option codes will become invalid.<br />

CORINA (<strong>CardioSoft</strong> V6.0 only supports CORINA models 101 118 3x)<br />

CORINA is powered from a special AC power adapter. A second version<br />

of the acquisition module is available for use with the electrode<br />

application system.<br />

1. Turn off the PC and connect the acquisition module to the parallel<br />

port of your PC.<br />

2. Connect the cable of the AC power adapter.<br />

3. Connect the patient cable to the acquisition module.<br />

4. If you purchased an acquisition module that supports the electrode<br />

application system, connect the pneumatic lead to the acquisition<br />

module. (Refer to the Operator's Manual of the Electrode Application<br />

System.)<br />

Note<br />

The ECG signal input is isolated and defibrillation-proof if used<br />

in conjunction with the CORINA acquisition module (type CF).<br />

2-6 <strong>CardioSoft</strong> Revision A<br />