Miscellaneous: Troubleshooting<br />

Problem<br />

No communication<br />

with MUSE.<br />

The ECG grid is not<br />

visible in the PDF<br />

file.<br />

Message Not<br />

enough free<br />

memory on hard<br />

drive!<br />

After connection of<br />

the CAM-USB<br />

interface box to the<br />

PC, no ECGs are<br />

displayed<br />

Patient record<br />

cannot be opened<br />

and is reported to<br />

be locked.<br />

Remedy<br />

Incorrect network address of the MUSE server or shared<br />

directory (section “MUSE Tab” on page 11-55).<br />

Required options Request MUSE Data or Send Data to MUSE<br />

not enabled (section “MUSE Tab” on page 11-55).<br />

Incorrect network configuration (system administrator).<br />

Incorrect Internet Explorer setup (version >4.0 is required, also<br />

for sending data to MUSE).<br />

Set up Acrobat Distiller as follows<br />

Acrobat V6.0:<br />

Start -> Settings -> Printer -> Adobe PDF -> menu Printer -><br />

Document Defaults -> menu Adobe PDF Settings ->disable do<br />

not send fonts to distiller;<br />

menu Paper/Quality -> button Advanced -> Graphic and select<br />

the field Download True Type Font as Softfont.<br />

Acrobat V5.0:<br />

Start -> Settings -> Printer -> Acrobat Distiller -> menu Printer -><br />

Document Defaults -> menu Adobe PDF Settings ->disable do<br />

not send fonts to distiller;<br />

menu Paper/Quality -> button Advanced -> Graphic and select<br />

the field Download True Type Font as Softfont .<br />

Acrobat V4.0<br />

Start -> Programs -> Adobe -> Acrobat 4.0 -> Acrobat Distiller<br />

4.0 -> Defaults -> Settings -> Fonts enable -> Embed all fonts.<br />

Free storage capacity on the hard drive where your database is<br />

stored, e.g., by compressing exercise test data or by deleting<br />

examinations. Also ensure that there is enough free storage<br />

capacity on the hard drive where your operating system is<br />

located and in the temporary directories (100 MB minimum).<br />

Remove connector of the CAM-USB interface box from the PC,<br />

wait about 10 seconds, then reconnect. When prompted to install<br />

a new USB device, clear the prompt or any other message with<br />

Cancel.<br />

In a network environment, the currently open patient record may<br />

be locked following the crash of the client or server (e.g. due to<br />

power failure). Please contact Customer Service for advice.<br />

Note<br />

Disable the SpeedStep technology when running the program on a<br />

laptop.<br />

B-28 <strong>CardioSoft</strong> Revision A<br />