Miscellaneous: Application Tips<br />

If other electrodes are used, we recommend disconnecting the patient<br />

cable from the system before delivering the defibrillation shock.<br />

Cubic Spline and the FRF algorithm cause a signal delay of approx. 2<br />

seconds; therefore they should be disabled if the patient has to be<br />

defibrillated while the ECG is being recorded (“Acquisition Tab” on page<br />

11-3).<br />

We recommend using non-polarizing disposable electrodes with<br />

defibrillation recovery ratings as specified in AAMI EC 12.3.2.2.4 (MMS<br />

P/N 900703 Silver Trace). AAMI EC12 requires that the polarization<br />

potential of an electrode pair does not exceed 100 mV 5 seconds after a<br />

defibrillation discharge.<br />

Note<br />

If excessive DC voltages are present at the electrode, then a message<br />

will appear indicating a Lead Off condition.<br />