Miscellaneous: Connecting Peripheral Devices<br />

Putting the Lode Excalibur Ergometer Into Service<br />

Every time you turn on the ergometer, the TERMINAL option must be<br />

selected in the ergometer menu and confirmed with Enter. Then the<br />

display will show the workload and pedal speed. THEN start the system.<br />

Otherwise an error message will appear, saying "No communication with<br />

ergometer/treadmill".<br />

Putting the EC 1200 Ergometer Into Service<br />

The EC 1200 bicycle ergometer only supports graded phases. When first<br />

activating the Protocol Editor, a single line will be displayed for<br />

<br />

<br />

<br />

<br />

<br />

<br />

Basic Load<br />

Load Increase<br />

Stage Time<br />

Number of Stages<br />

ECG Recording Interval, and<br />

BP Interval<br />

for each phase.<br />

Basic Load is the load level at the beginning of a phase. When the Stage<br />

Time has elapsed, the next stage begins and the load increases by the<br />

value entered at Load Increase.<br />

At Number of Stages you enter the number of stages for a phase. The<br />

configured ECG Recording Interval and BP Interval ensure that ECG<br />

recordings and BP measurements are initiated at the appropriate point<br />

in time before the end of a stage. The first ECG recording and BP<br />

measurement are fixed and cannot be changed. Enter a minimum value<br />

of 10 seconds for the ECG Recording Interval and of 1 minute for the BP<br />

Interval. The ECG Recording Interval and the BP Interval will be<br />

maintained until the end of the phase. These values can thus be set<br />

independently of the Stage Time. Each time the program advances to the<br />

next stage both intervals are reset to synchronize them with the new<br />

stage.<br />

A BP Interval must be entered in the pretest phase to take a blood<br />

pressure measurement before the test. Since no further BP<br />

measurements are required in pretest, the default BP Interval is 99<br />

minutes. If you do not want to take a pretest BP measurement, change<br />

this value to 00:00:00.<br />

A BP monitor connected directly to the EC 1200 will take the blood<br />

pressure at the beginning of each phase.<br />

The arrow buttons W+ and W– change the load setting to the value of the<br />

next or previous stage. All other protocol parameters remain unchanged.<br />

The load for subsequent stages increases accordingly.<br />