Miscellaneous: Connecting Peripheral Devices<br />

Treadmills<br />

Connect the treadmills to the COM 1 or COM 2 port.<br />

Model<br />

Treadmills without digital communication<br />

interface<br />

GE <strong>Medical</strong> Systems Information<br />

Technologies treadmills<br />

1800<br />

1900<br />

T2000/T2100<br />

T2100<br />

Trackmaster TMX 425<br />

Trackmaster TM400<br />

internal treadmill<br />

Series 2000<br />

T2100<br />

TMX 425<br />

TM 400<br />

Configuration<br />

Note<br />

Other bicycle ergometers and treadmill models with an analog port<br />

can be connected to the analog output of the system.<br />

Connecting the ERGOLINE 900/900 L/900 EL Ergometers to the System<br />

Set the following parameters at the ergometer:<br />

ECG type: P10<br />

Baud rate: 4800<br />

The ergometer software must be version V436 or V236; ergometers with<br />

a 4-button control panel must be software version 752 or M343.<br />

Plug the round connector of the connecting cable into the J10 / RS 232<br />

port of the ergometer and the other end into the COM 1 or COM 2 port of<br />

the system.<br />

B-4 <strong>CardioSoft</strong> Revision A<br />

2023324-053