System Settings: Exercise Test Setup<br />

TWA Tab<br />

Click the TWA tab to open the window.<br />

During the exercise test the TWA algorithm analyzes the data in the<br />

background. In each of the selected leads, the algorithm calculates the<br />

even and odd medians, the TWA value and the noise limit. The data are<br />

saved at intervals of 15 minutes. Verify the TWA settings on this tab<br />

before the test.<br />

A B C D<br />

E<br />

923A-S<br />

A<br />

B<br />

C<br />

D<br />

E<br />

Select the TWA Heart Rate Limit (TWA values whose HR exceeds this limit<br />

are excluded from calculation of the maximum TWA value and are identified<br />

with a question mark on the display).<br />

Select the TWA Noise Limit (TWA values whose noise value exceeds this limit<br />

are excluded from calculation of the maximum TWA value and are identified<br />

with a question mark on the display).<br />

Heart rate: when first reached, the corresponding TWA value is stored<br />

separately.<br />

Click to reset the values to the defaults.<br />

Click down arrow to select the update factor.<br />

Note<br />

During the test, TWA data records are saved at given points in time.<br />

Each TWA data record provides the following information:<br />

<br />

<br />

<br />

time it was saved<br />

heart rate<br />

TWA and noise values for each lead<br />

Revision A <strong>CardioSoft</strong> 11-31<br />

2023324-053