System Settings: Exercise Test Setup<br />

I<br />

J<br />

K<br />

Select the Max. Load Interpolation function (only possible when the exercise test<br />

is performed on a bicycle ergometer). The maximum load is the load of the last<br />

completed stage plus the duration of the last stage x load increase/stage time.<br />

Example:<br />

load of the last completed stage: 125 Watts<br />

last stage aborted after 1 minute<br />

load increase: 25 Watts<br />

stage time: 2 minutes<br />

125 W + 1 x 25/2 = 137.5 W<br />

This value is also used to determine the percentage of the target load.<br />

Select to Save full disclosure ECG, then choose the phase when storage of the<br />

full disclosure ECG should start, and the sampling rate.<br />

Select the METS Interpolation Time.<br />

Revision A <strong>CardioSoft</strong> 11-25<br />