Exercise Test: During the Test<br />

Manual Control<br />

F11: comment Press to enter comments to appear in the Tabular<br />

Summary.<br />

F12: enter BP Press to enter BP readings.<br />

12 lead Press to start an ECG recording (12 leads, 5 seconds<br />

from memory, 5 seconds of real-time data).<br />

recall<br />

Press to print a 10-second delayed recall report.<br />

rhythm<br />

Press to start a continuous rhythm report.<br />

stop writer Press to stop the writer/printer.<br />

medians Press to print a medians report.<br />

arrhy doc<br />

Press to enable or disable the automatic<br />

documentation of arrhythmias.<br />

Note<br />

Cubic Spline / FRF algorithms can delay the rhythm report by<br />

approximately 1 second when enabled.<br />

12SL<br />

F7: grade +<br />

F8: grade -<br />

F5: speed +<br />

or load +<br />

F6: speed -<br />

or load -<br />

hold stage<br />

Press to acquire a 12SL resting ECG with<br />

measurements and interpretation (only in the pretest<br />

and recovery phases).<br />

Press to change the elevation of the treadmill belt.<br />

Press to change the belt speed or ergometer load.<br />

Press to hold the exercise test in the current stage.<br />

Note<br />

<br />

<br />

<br />

<br />

In the pretest phase, Shift + pretest or Shift + F1 allow you to<br />

return to the previous stage.<br />

If a bicycle ergometer test is initiated with the F2: exercise<br />

button, the clock will start even when the patient has not yet<br />

begun to pedal.<br />

When the patient reaches the target HR, the HR reading turns<br />

red and an audio signal sounds (if configured, see section<br />

“Miscellaneous 1 Tab” on page 11-24).<br />

Using the treadmill speed and grade buttons during a protocol<br />

puts the treadmill in manual control. The system is no longer<br />

controlled automatically by the protocol. The system will enter a<br />

new stage and the phase name appearing in the Tabular<br />

Summary will be Manual.<br />

Revision A <strong>CardioSoft</strong> 6-15<br />

2023324-053