Exercise Test: Before the Test<br />

Operating Steps<br />

1. Check the applied electrodes and verify the signal quality in all<br />

leads.<br />

ST measurements and arrhythmia analysis will automatically be<br />

displayed when all necessary electrodes have been applied and<br />

signals of adequate quality are obtained (e.g. without major baseline<br />

fluctuations). If all electrodes are not being used, click Relearn<br />

Medians to manually activate ST measurement and arrhythmia<br />

analysis.<br />

2. Click Select Protocol to verify that the appropriate test protocol is<br />

active or to select another protocol.<br />

3. Click Custom Setup to select and load user-specific test settings.<br />

4. Click Setup to edit the exercise test setup.<br />

Note<br />

Once you initiate the pretest phase you can no longer change the<br />

test settings.<br />

5. Press F1: pretest to initiate the test. The pretest screen will appear<br />

and the pretest phase begins.<br />

Revision A <strong>CardioSoft</strong> 6-11<br />

2023324-053