<strong>CombiFlash</strong> Torrent® <strong>User</strong> <strong>Manual</strong><br />

PeakTrak’s default Run Length units by changing the<br />

CONFIGURATION [4.3.10] settings.<br />

4.3.2.2 Run Settings<br />

• RediSep Column — The RediSep column media and<br />

size are detected by PeakTrak and reported here. The<br />

column should match the one that will be used for the<br />

run. You may override the detected column size by<br />

selecting a different size. See also: REDISEP RF COLUMN<br />

SELECTION GUIDE [ON-LINE] and ABOUT REDISEP RF<br />

COLUMN DETECTION [ON-LINE].<br />

• Sample Name — The sample name is a text entry box<br />

in which you can label the run. Type a unique descriptor.<br />

PeakTrak saves the run information under this name.<br />

Therefore, PeakTrak cannot accept any characters that<br />

are reserved by the operating system (* ? / \, etc.). If you<br />

do not type a Sample Name, PeakTrak will generate a<br />

date/time stamp for the name when you click the Play<br />

button to begin the run.<br />

• Flow Rate — Type or select the desired flow rate for the<br />

run. When using the <strong>Isco</strong> default column methods, the<br />

module sets the flow rate to the optimum flow rate of the<br />

column.<br />

• Equilibration Volume — Type or select the volume of<br />

solvent that is pumped through the column and out the<br />

waste port before the sample is introduced. Data is not<br />

collected while this volume is being pumped. Three or<br />

more column volumes are recommended for most<br />

applications.<br />

• Initial Waste & Time Windows — Click this button to<br />

open a window from which you can view and modify<br />

these settings. For the Initial Waste, type or select the<br />

volume of eluate that is allowed to pass through the<br />

column to the waste port after the sample is injected.<br />

After this volume has been diverted, the module is ready<br />

to collect peak or all fluids in the collection tubes. Use<br />

this feature to clear the internal lines of any elute left<br />

from a previous run and to conserve collection tubes by<br />

diverting all fluids until the first peak is expected.<br />

4-16