<strong>CombiFlash</strong> Torrent® <strong>User</strong> <strong>Manual</strong><br />

• Sample Name — The sample name is a text entry box<br />

in which you can label the run. Type a unique descriptor.<br />

PeakTrak saves the run information under this name.<br />

Therefore, PeakTrak cannot accept any characters that<br />

are reserved by the operating system (* ? / \, etc.). If you<br />

do not type a Sample Name, PeakTrak will generate a<br />

date/time stamp for the name when you click the Play<br />

button to begin the run.<br />

• Flow Rate — Type or select the desired flow rate for the<br />

run. When using the <strong>Isco</strong> default column methods,<br />

PeakTrak sets the flow rate to the optimum flow rate of<br />

the column.<br />

• Detector — Reports the current peak detection mode.<br />

To change the peak detection settings, open the METHOD<br />

EDITOR [4.3.2] window.<br />

• Solvent A — Select the desired solvent from the list.<br />

The solvents in the list are those defined by the<br />

CONFIGURATION [4.3.10] window. For assistance in<br />

selecting a solvent, see the SOLVENT SELECTION TABLE<br />

[ON-LINE].<br />

• Solvent B — Select the desired solvent from the list.<br />

4.3.1.3 Run Control Buttons<br />

From left to right, the control buttons are:<br />

• Play — The Play button starts or resumes the run. After<br />

a run has started, this button is replaced by the Pause<br />

button.<br />

• Pause — The Pause button holds the %B at the current<br />

value while the module continues to operate (sometimes<br />

called an isocratic hold). Note that pausing the run<br />

extends the run length. While in the Paused state, you<br />

can resume the run by clicking the Play button, or stop<br />

the run by clicking the Stop button. If you resume the<br />

run, the module continues the gradient curve from the<br />

%B when the module was paused.<br />

• Stop — This button suspends the entire run. Unlike the<br />

Paused state, the pump, peak detection, and fraction<br />

collection will not operate. While stopped, you can save<br />

4-12