CombiFlash Torrent® User Manual - Isco
CombiFlash Torrent® User Manual - Isco
CombiFlash Torrent® User Manual - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>CombiFlash</strong> Torrent® <strong>User</strong> <strong>Manual</strong><br />
• Sample Name — The sample name is a text entry box<br />
in which you can label the run. Type a unique descriptor.<br />
PeakTrak saves the run information under this name.<br />
Therefore, PeakTrak cannot accept any characters that<br />
are reserved by the operating system (* ? / \, etc.). If you<br />
do not type a Sample Name, PeakTrak will generate a<br />
date/time stamp for the name when you click the Play<br />
button to begin the run.<br />
• Flow Rate — Type or select the desired flow rate for the<br />
run. When using the <strong>Isco</strong> default column methods,<br />
PeakTrak sets the flow rate to the optimum flow rate of<br />
the column.<br />
• Detector — Reports the current peak detection mode.<br />
To change the peak detection settings, open the METHOD<br />
EDITOR [4.3.2] window.<br />
• Solvent A — Select the desired solvent from the list.<br />
The solvents in the list are those defined by the<br />
CONFIGURATION [4.3.10] window. For assistance in<br />
selecting a solvent, see the SOLVENT SELECTION TABLE<br />
[ON-LINE].<br />
• Solvent B — Select the desired solvent from the list.<br />
4.3.1.3 Run Control Buttons<br />
From left to right, the control buttons are:<br />
• Play — The Play button starts or resumes the run. After<br />
a run has started, this button is replaced by the Pause<br />
button.<br />
• Pause — The Pause button holds the %B at the current<br />
value while the module continues to operate (sometimes<br />
called an isocratic hold). Note that pausing the run<br />
extends the run length. While in the Paused state, you<br />
can resume the run by clicking the Play button, or stop<br />
the run by clicking the Stop button. If you resume the<br />
run, the module continues the gradient curve from the<br />
%B when the module was paused.<br />
• Stop — This button suspends the entire run. Unlike the<br />
Paused state, the pump, peak detection, and fraction<br />
collection will not operate. While stopped, you can save<br />
4-12