Section 4 <strong>User</strong> Help Reference<br />
the left side of the window changes to show a collection point<br />
map or the method parameters in a PEAK COLLECTION DATA<br />
[4.3.1.5] pane.<br />
4.3.1.1 Gradient Plot<br />
The gradient plot area depicts the current gradient that will<br />
be used for the run. The X-axis of the plot displays the run<br />
time in minutes or COLUMN VOLUMES [ON-LINE]. The X-axis scale<br />
can be adjusted by typing or selecting the Run Length. The<br />
left Y-axis displays absorbance units and the right Y-axis<br />
indicates the percentage of solvent B. By default, PeakTrak<br />
automatically scales the left Y-axis to best display the<br />
absorbance trace. PeakTrak also sets the right Y-axis to 100%.<br />
You can override the Y-axes scales by clicking the Format<br />
Graph button and setting the desired values.<br />
The Cartridge Bypass button may be used during a run to<br />
relieve backpressure caused by impurities precipitating in the<br />
solid load cartridge. Use this feature with care to avoid leaving<br />
desired compound on the cartridge. Read more about this<br />
feature in BYPASSING THE SOLID LOAD CARTRIDGE [4.4.8]<br />
If the external detector option is enabled, the plot area will<br />
include a secondary left Y-axis to display the analog input<br />
trace.<br />
The gradient plot area also provides a convenient method to<br />
edit the gradient curve. You can click and drag any of the<br />
points to change the shape of the curve and use the buttons at<br />
the bottom of the plot area to insert and delete points. For<br />
complete instructions on defining gradient curves, see<br />
DEFINING A GRADIENT [4.4.5].<br />
4.3.1.2 Run Settings<br />
• RediSep — PeakTrak reports the detected RediSep<br />
column media and size. The column should match the<br />
one that will be used for the run. You may override the<br />
detected column size by selecting a different size or<br />
media. See also: REDISEP RF COLUMN SELECTION GUIDE<br />
[ON-LINE] and ABOUT REDISEP RF COLUMN DETECTION<br />
[ON-LINE].<br />
4-11