Section 3 Operation<br />

Note<br />

When you click the Next Tube button, the system immediately<br />

marks the tube advance on the chromatogram. However, the<br />

system delays the movement of the drop former over the tubes.<br />

This delay accounts for the volume held in the tubing between the<br />

detector and the drop former, sometimes called the “delay<br />

volume.” It is important to remember that the chromatogram<br />

displays activity in real time while the fraction collector activity lags<br />

behind. The duration of this delay will vary with the flow rate.<br />

• Rewind — This button is active when the system has<br />

completed the run, or if the run was stopped by clicking<br />

the Stop button. The Rewind button returns you to the<br />

Main window with the current method.<br />

If you have stopped the run before its programmed run<br />

length has elapsed, clicking the Rewind button changes the<br />

current method to the new, shorter run length. This modified<br />

method is ready for the next run, or can be saved for<br />

future runs.<br />

• Fast Forward — Click this button to jump to the next<br />

step of a run.<br />

Once you have started a run, the system performs several<br />

steps. The first step is to deliver solvents using the programmed<br />

gradient for the entire run length. When this step<br />

is complete, the system will purge the column with air and<br />

clean the injection valve. Clicking the Fast Forward button<br />

will cause the system to skip any remaining time in the<br />

current step and advance to the beginning of the next step.<br />

Unlike the Rewind button, clicking the Fast Forward button<br />

to complete the run will not modify the current<br />

method’s run length setting.<br />

Note<br />

Solvent may be under pressure. Do not remove a column or solid<br />

load cartridge until the pressure gauge on the touch screen to<br />

return to 0 psi.<br />

3-17