• If you selected Solid (Pause), the system<br />
equilibrates the column with the starting %B solvent<br />
mixture and then waits while you prepare the solid<br />
sample load cartridge. When prompted, place the<br />
cartridge on the system and click OK to continue.<br />
• If you selected Solid, the system equilibrates the<br />
column and immediately starts the separation. The<br />
prepared solid load cartridge should be in place<br />
before clicking the OK button.<br />
• If you selected None (on Column), the system skips<br />
the equilibration and immediately starts the<br />
separation. The sample must be on the column before<br />
clicking the OK button.<br />
• If you selected Liquid, the system equilibrates the<br />
column and then waits. When instructed, load the<br />
liquid sample using the syringe injection (3.4.1) or the<br />
Sample Load Pump (3.4.2) options.<br />
3.4.1 Syringe Injection<br />
Refer to Figure 3-6 for the following steps.<br />
1. Connect the Liquid Injection Adapter from the accessory<br />
kit to the Sample Injection Port.<br />
2. Place the syringe on top of the Liquid Injection Adapter<br />
valve.<br />
Closed: handle<br />
perpendicular<br />
to tubing<br />
Open: handle<br />
in-line with tubing<br />
Figure 3-6 Liquid Injection Adapter<br />
3-12