• If you selected Solid (Pause), the system<br />

equilibrates the column with the starting %B solvent<br />

mixture and then waits while you prepare the solid<br />

sample load cartridge. When prompted, place the<br />

cartridge on the system and click OK to continue.<br />

• If you selected Solid, the system equilibrates the<br />

column and immediately starts the separation. The<br />

prepared solid load cartridge should be in place<br />

before clicking the OK button.<br />

• If you selected None (on Column), the system skips<br />

the equilibration and immediately starts the<br />

separation. The sample must be on the column before<br />

clicking the OK button.<br />

• If you selected Liquid, the system equilibrates the<br />

column and then waits. When instructed, load the<br />

liquid sample using the syringe injection (3.4.1) or the<br />

Sample Load Pump (3.4.2) options.<br />

3.4.1 Syringe Injection<br />

Refer to Figure 3-6 for the following steps.<br />

1. Connect the Liquid Injection Adapter from the accessory<br />

kit to the Sample Injection Port.<br />

2. Place the syringe on top of the Liquid Injection Adapter<br />

valve.<br />

Closed: handle<br />

perpendicular<br />

to tubing<br />

Open: handle<br />

in-line with tubing<br />

Figure 3-6 Liquid Injection Adapter<br />

3-12