Section 3 Operation<br />

cartridge. After you have placed the cartridge on the<br />

system, click OK to continue with the run.<br />

• Solid – Select this option if you already have<br />

prepared the sample and placed it into a solid sample<br />

load cartridge. The system will proceed automatically<br />

until the end of the programmed run.<br />

• Liquid – Select this option if you have prepared a<br />

liquid sample and plan to manually inject it or pump<br />

into the injection port after column equilibration.<br />

• None (on column) – Select this option if you have<br />

preloaded the sample on the column. The system will<br />

skip column equilibration so that the sample will not<br />

be flushed from the column before the run.<br />

5. Review or update the Start Rack and Start Tube settings.<br />

6. Review the Minimum Run Requirements. The system will<br />

report the estimated solvent volumes, expected waste,<br />

collection tube usage, etc. You can use this information to<br />

verify that there will be enough solvent to complete the<br />

run, the waste volume does not exceed the collection<br />

container’s capacity, and whether or not more collection<br />

tubes will be required during the run.<br />

Note<br />

The Torrent estimates the solvent volumes by monitoring the<br />

solvent level in the container and the known usage rate. The<br />

systems continue to refine this estimate during operation.<br />

Note<br />

If the waste level sensing tube is inserted correctly into the waste<br />

container, the Torrent will automatically suspend operation before<br />

an overflow condition might exist. To prevent the run from being<br />

suspended before completing the run, ensure that the container<br />

will hold the expected waste volume.<br />

7. Click OK and the run begins. The system responds<br />

according to the sample Loading Type you selected in<br />

step 4.<br />

3-11