Section 3 Operation<br />
cartridge. After you have placed the cartridge on the<br />
system, click OK to continue with the run.<br />
• Solid – Select this option if you already have<br />
prepared the sample and placed it into a solid sample<br />
load cartridge. The system will proceed automatically<br />
until the end of the programmed run.<br />
• Liquid – Select this option if you have prepared a<br />
liquid sample and plan to manually inject it or pump<br />
into the injection port after column equilibration.<br />
• None (on column) – Select this option if you have<br />
preloaded the sample on the column. The system will<br />
skip column equilibration so that the sample will not<br />
be flushed from the column before the run.<br />
5. Review or update the Start Rack and Start Tube settings.<br />
6. Review the Minimum Run Requirements. The system will<br />
report the estimated solvent volumes, expected waste,<br />
collection tube usage, etc. You can use this information to<br />
verify that there will be enough solvent to complete the<br />
run, the waste volume does not exceed the collection<br />
container’s capacity, and whether or not more collection<br />
tubes will be required during the run.<br />
Note<br />
The Torrent estimates the solvent volumes by monitoring the<br />
solvent level in the container and the known usage rate. The<br />
systems continue to refine this estimate during operation.<br />
Note<br />
If the waste level sensing tube is inserted correctly into the waste<br />
container, the Torrent will automatically suspend operation before<br />
an overflow condition might exist. To prevent the run from being<br />
suspended before completing the run, ensure that the container<br />
will hold the expected waste volume.<br />
7. Click OK and the run begins. The system responds<br />
according to the sample Loading Type you selected in<br />
step 4.<br />
3-11