<strong>CombiFlash</strong> Torrent® <strong>User</strong> <strong>Manual</strong><br />

Note<br />

After loading a RediSep Rf column, the system will use RFID<br />

technology to automatically detect the media type and column<br />

size. PeakTrak displays the detected column size on the Main and<br />

Method Editor windows. If the system does not detect the column,<br />

manually select the column media and size.<br />

3.4 Start a Default Method<br />

After completing the system installation steps (Section 2),<br />

preparing the sample, and inserting the column, you are ready<br />

to perform a run with the default method. The Torrent’s<br />

default methods are factory-set with run parameters typically<br />

used by chemists. Default methods are optimized for RediSep<br />

Rf columns. The default settings will:<br />

• collect all fluid in the fraction collector rack<br />

• differentiate between UV (254 nm) vs. non-UV absorbing<br />

fluid and place the fluids into separate tubes<br />

• cut UV absorbing peaks based on slope detection or level<br />

threshold.<br />

To start the run from the PeakTrak Main window:<br />

1. Enter a sample name. If you choose not to, PeakTrak will<br />

enter the date and time as the sample name when you start<br />

the run.<br />

2. Review the PeakTrak Main window settings. If you want to<br />

change any settings, refer to the Editing a Method on-line<br />

help topic.<br />

3. Click the Play button.<br />

4. Select the sample Loading Type from the list. This Loading<br />

Type should support the sample preparation you chose in<br />

Section 3.2. Possible sample loading types are:<br />

• Solid (Pause) – Select this option if you intend place<br />

the sample into a solid sample load cartridge, but<br />

have not yet prepared the sample. When you click the<br />

run button, the system will perform a column<br />

equilibration and then wait while you prepare the<br />

3-10