Note<br />
After loading a RediSep Rf column, the system will use RFID<br />
technology to automatically detect the media type and column<br />
size. PeakTrak displays the detected column size on the Main and<br />
Method Editor windows. If the system does not detect the column,<br />
manually select the column media and size.<br />
3.4 Start a Default Method<br />
After completing the system installation steps (Section 2),<br />
preparing the sample, and inserting the column, you are ready<br />
to perform a run with the default method. The Torrent’s<br />
default methods are factory-set with run parameters typically<br />
used by chemists. Default methods are optimized for RediSep<br />
Rf columns. The default settings will:<br />
• collect all fluid in the fraction collector rack<br />
• differentiate between UV (254 nm) vs. non-UV absorbing<br />
fluid and place the fluids into separate tubes<br />
• cut UV absorbing peaks based on slope detection or level<br />
threshold.<br />
To start the run from the PeakTrak Main window:<br />
1. Enter a sample name. If you choose not to, PeakTrak will<br />
enter the date and time as the sample name when you start<br />
the run.<br />
2. Review the PeakTrak Main window settings. If you want to<br />
change any settings, refer to the Editing a Method on-line<br />
help topic.<br />
3. Click the Play button.<br />
4. Select the sample Loading Type from the list. This Loading<br />
Type should support the sample preparation you chose in<br />
Section 3.2. Possible sample loading types are:<br />
• Solid (Pause) – Select this option if you intend place<br />
the sample into a solid sample load cartridge, but<br />
have not yet prepared the sample. When you click the<br />
run button, the system will perform a column<br />
equilibration and then wait while you prepare the<br />
3-10