<strong>CombiFlash</strong> Torrent® <strong>User</strong> <strong>Manual</strong><br />

The Direct Injection Adapter is recommended when using the<br />

Liquid Load Pump to load samples directly on the column, or<br />

when there is a need to bypass the automated injection valve<br />

during sample injections.<br />

3.1.4 Liquid Injection Adapter<br />

The Liquid Injection Adapter (60-5244-059) should be used<br />

when performing liquid sample injections with a syringe. The<br />

valve prevents the liquid from be forced back out of the<br />

injection port. Refer to section 3.4.1 for more information.<br />

3.1.5 Prime Tube Assembly<br />

The prime tube assembly (209-0165-76), also called a column<br />

bypass tube, can be used in place of a column when<br />

purging/priming the system and when performing<br />

maintenance.<br />

3.1.6 Prime Pump Plug, 1/8-inch<br />

Use this plug (209-0165-21) to cap the B Solvent’s Prime port<br />

when this port is not in use.<br />

3.1.7 Stylus<br />

The stylus with adhesive-backed mounting plate may be<br />

attached near the LCD display for precise control of the touch<br />

screen.<br />

3.1.8 Test Mix B Vials and 5 mL Syringe<br />

Test Mix B can be used to determine if the system is operating<br />

correctly. Refer to SYSTEM VERIFICATION WITH TEST MIX B in<br />

section 2.15 for instructions.<br />

3.1.9 Cable Clips and Ties<br />

Use the clips and ties to secure tubing or wiring after<br />

installation.<br />

3.2 Sample Preparation<br />

Before starting a run, consider how the sample will be<br />

introduced to the column media. This section discusses three<br />

methods: liquid injection, solid sample cartridges, and<br />

preloading the sample on the column.<br />

3-4