<strong>CombiFlash</strong> Torrent® <strong>User</strong> <strong>Manual</strong><br />

third extensions. At 24 minutes, the %B solvent strength<br />

returned to zero. Before the final minute elapsed, more<br />

compound was detected, causing PeakTrak to extend the run<br />

for a final five minutes at the minimum %B.<br />

Enable rapid equilibration – select this option to<br />

equilibrate the column at a high flow rate.<br />

Due to the high pressure that is possible during rapid<br />

equilibration, this option may not be desirable for some<br />

column media or purification methods.<br />

2.13.3 Fraction Collector<br />

Use this tab to configure the Torrent to control up to four Foxy<br />

R2 High Flow fraction collectors. This tab is also used to<br />

configure the Torrent for use with a Fractionation Valve.<br />

Foxy fraction collector(s) should be connected to the Torrent<br />

via a straight-through network cable. To communicate with<br />

each fraction collector, the IP address settings on this tab must<br />

match the IP address set in each fraction collector. Note that<br />

all IP addresses must be unique. Refer to the fraction collector<br />

user manual to set the address of each module.<br />

Note<br />

The optional Foxy R2 High Flow fraction collector does not support<br />

DHCP. If you are adding the Torrent module with optional Foxy R2<br />

High Flow fraction collectors to your organization’s network, a<br />

static IP address must be assigned to each fraction collector.<br />

Contact your IT department for assistance.<br />

A delay volume must also be specified for each Foxy connected<br />

as a system. If using the Fractionation Valve instead, a delay<br />

volume should also be entered. The delay volume entries<br />

ensure that the tube changes occur when the peak fluids are<br />

present at the fraction collector diverter valve.<br />

If using the factory-cut tubing provided with the Foxy R2 High<br />

Flow fraction collectors or Fractionation Valve, enter the<br />

following as applicable:<br />

Foxy 1 Delay Volume – 20 mL<br />

Foxy 2 Delay Volume – 40 mL<br />

2-34