third extensions. At 24 minutes, the %B solvent strength<br />
returned to zero. Before the final minute elapsed, more<br />
compound was detected, causing PeakTrak to extend the run<br />
for a final five minutes at the minimum %B.<br />
Enable rapid equilibration – select this option to<br />
equilibrate the column at a high flow rate.<br />
Due to the high pressure that is possible during rapid<br />
equilibration, this option may not be desirable for some<br />
column media or purification methods.<br />
2.13.3 Fraction Collector<br />
Use this tab to configure the Torrent to control up to four Foxy<br />
R2 High Flow fraction collectors. This tab is also used to<br />
configure the Torrent for use with a Fractionation Valve.<br />
Foxy fraction collector(s) should be connected to the Torrent<br />
via a straight-through network cable. To communicate with<br />
each fraction collector, the IP address settings on this tab must<br />
match the IP address set in each fraction collector. Note that<br />
all IP addresses must be unique. Refer to the fraction collector<br />
user manual to set the address of each module.<br />
Note<br />
The optional Foxy R2 High Flow fraction collector does not support<br />
DHCP. If you are adding the Torrent module with optional Foxy R2<br />
High Flow fraction collectors to your organization’s network, a<br />
static IP address must be assigned to each fraction collector.<br />
Contact your IT department for assistance.<br />
A delay volume must also be specified for each Foxy connected<br />
as a system. If using the Fractionation Valve instead, a delay<br />
volume should also be entered. The delay volume entries<br />
ensure that the tube changes occur when the peak fluids are<br />
present at the fraction collector diverter valve.<br />
If using the factory-cut tubing provided with the Foxy R2 High<br />
Flow fraction collectors or Fractionation Valve, enter the<br />
following as applicable:<br />
Foxy 1 Delay Volume – 20 mL<br />
Foxy 2 Delay Volume – 40 mL<br />
2-34