Automatically Print Report at End of Run – if this option<br />
is enabled, the system will print a report at the end of each<br />
run.<br />
Note<br />
This option requires the system to be configured for network<br />
operation and a connection to a printer on the network. Network<br />
settings are discussed in Section 5, REMOTE INTERFACES. Do not<br />
select this option for these initial installation steps.<br />
Beeper – If enabled, the Torrent will sound an alert whenever<br />
user intervention is required during a run.<br />
Default Run Units – Run units are displayed along the<br />
X-axis of the chromatogram. Select Time (in minutes) or<br />
Column Volumes. A column volume unit is the time it takes to<br />
pump enough solvent to exchange the volume held by the<br />
column. The duration in minutes will vary according to the<br />
column media and size, and the flow rate.<br />
Gradient Method – Select the standard gradient type<br />
(Figure 2-20).<br />
Linear<br />
Figure 2-20 Gradient types<br />
Step<br />
Isocratic<br />
• Linear – The system gradually mixes solvents A and B<br />
from the last programmed %B until it reaches the next<br />
programmed %B. This method connects the inflection<br />
points that define the gradient with a straight line.<br />
• Step – The system holds each %B value until it reaches<br />
the next programmed %B. Because the system holds the<br />
%B values, the changes at each new value are nearly<br />
instantaneous, producing a stepped appearance on the<br />
plot area.<br />
• Isocratic – The system holds the initial solvent mix for<br />
the entire run.<br />
2-32