<strong>CombiFlash</strong> Torrent® <strong>User</strong> <strong>Manual</strong><br />

level sensing feature (recommended). Otherwise, uncheck this<br />

option to disable this feature.<br />

Solvent B Priming Pump – Select the “Enable solvent B<br />

priming pump” check box to enable this feature<br />

(recommended). Clear this box to disable this feature.<br />

The Torrent is equipped with a priming pump for the B<br />

solvent to ensure that the system forms repeatable gradients.<br />

This pump can be enabled or disabled.<br />

Note<br />

If the Torrent is connected to pressurized lines, disable the solvent<br />

B priming pump.<br />

<strong>CombiFlash</strong> Torrent Name – (Optional) Use this option to<br />

name your system. The name will appear in operational<br />

displays and run summaries. This feature is useful when your<br />

laboratory has more than one Torrent.<br />

Time Zone – Select your time zone from the drop-down list<br />

box.<br />

Set Date/Time – Click this button to open a window from<br />

which you can set the system date and time.<br />

To prevent unauthorized time changes, this feature is<br />

password protected. The system is shipped with the password<br />

set to combiflash. Use the Tools>Set Password menu<br />

command to change this password for greater security.<br />

Vapor Limit – The Torrent has an internal vapor sensor that<br />

detects vapors present in the Torrent. When the limit is<br />

exceeded, the Torrent will shut down to avoid a hazardous<br />

condition.<br />

Teledyne <strong>Isco</strong> recommends using the default setting of 12.<br />

This value represents an organic vapor level slightly above the<br />

ambient vapor level of a well-ventilated laboratory.<br />

A setting below 12 could cause random alarms without<br />

significant vapors present. If PeakTrak displays a Vapor Limit<br />

alarm while the Vapor Limit is set at 12, perform the following<br />

checks on your laboratory and the instrument:<br />