Section 2 Preparation<br />
(sections 2.13.1 through 2.13.3) for the desired operation and<br />
then click OK to save the settings.<br />
Some configuration settings do not take effect until the system<br />
is restarted. After completing sections 2.13.1 through 2.13.3<br />
follow the instructions in section 2.13.4.<br />
2.13.1 Instrument Configuration<br />
Solvents – When you installed the system, two solvents were<br />
connected to the back panel. This section of the Instrument<br />
Configuration tab allows you to name these solvents. To do so:<br />
1. Click the “Add Solvent” button.<br />
2. Enter the solvent name and click the OK button.<br />
3. Repeat steps 1 and 2 for all solvents.<br />
Solvent Level Sensing – Select the “Enable solvent level<br />
sensing” check box to enable this feature (recommended).<br />
Clear this box to disable this feature.<br />
The solvent level sensing feature will monitor solvent usage to<br />
minimize the risk of running out of solvent during a<br />
purification run. The system compares the volume required<br />
for a purification run with the estimated volume in the solvent<br />
container and alerts you when there is not enough.<br />
Note<br />
If the Torrent is connected to pressurized lines, disable solvent<br />
level sensing.<br />
If using the solvent level sensing feature, enter the Minimum<br />
Solvent Level (density dependent) as a range in centimeters.<br />
To understand this range, consider that dichloromethane is<br />
more dense than hexane. If the selected range is 1–4, the<br />
system triggers an alert when the dichloromethane level rises<br />
to about 1 cm above the tip of the level sense line. At the same<br />
selected range, the system would trigger an alert when the<br />
hexane rises to about 4 cm above the tip of the level sense line.<br />
Waste Level Sensing – The Torrent can detect full container<br />
conditions if the air line is placed in the waste collection<br />
container. If waste level detection is desired, enable the waste<br />
2-29