07.03.2014 Views

CombiFlash Torrent® User Manual - Isco

CombiFlash Torrent® User Manual - Isco

CombiFlash Torrent® User Manual - Isco

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Section 2 Preparation<br />

(sections 2.13.1 through 2.13.3) for the desired operation and<br />

then click OK to save the settings.<br />

Some configuration settings do not take effect until the system<br />

is restarted. After completing sections 2.13.1 through 2.13.3<br />

follow the instructions in section 2.13.4.<br />

2.13.1 Instrument Configuration<br />

Solvents – When you installed the system, two solvents were<br />

connected to the back panel. This section of the Instrument<br />

Configuration tab allows you to name these solvents. To do so:<br />

1. Click the “Add Solvent” button.<br />

2. Enter the solvent name and click the OK button.<br />

3. Repeat steps 1 and 2 for all solvents.<br />

Solvent Level Sensing – Select the “Enable solvent level<br />

sensing” check box to enable this feature (recommended).<br />

Clear this box to disable this feature.<br />

The solvent level sensing feature will monitor solvent usage to<br />

minimize the risk of running out of solvent during a<br />

purification run. The system compares the volume required<br />

for a purification run with the estimated volume in the solvent<br />

container and alerts you when there is not enough.<br />

Note<br />

If the Torrent is connected to pressurized lines, disable solvent<br />

level sensing.<br />

If using the solvent level sensing feature, enter the Minimum<br />

Solvent Level (density dependent) as a range in centimeters.<br />

To understand this range, consider that dichloromethane is<br />

more dense than hexane. If the selected range is 1–4, the<br />

system triggers an alert when the dichloromethane level rises<br />

to about 1 cm above the tip of the level sense line. At the same<br />

selected range, the system would trigger an alert when the<br />

hexane rises to about 4 cm above the tip of the level sense line.<br />

Waste Level Sensing – The Torrent can detect full container<br />

conditions if the air line is placed in the waste collection<br />

container. If waste level detection is desired, enable the waste<br />

2-29