Section 2 Preparation<br />
If an optional waste cap is not used, ensure the waste lines are<br />
secured so they drain into the waste container. Secure the<br />
Waste Level Sense air tubing so its outlet is at least two inches<br />
(5 cm) below the container’s maximum level. Seal the<br />
container opening to avoid solvent vapors.<br />
2.8 Optional External Gas<br />
The Torrent has an internal air pump that provides<br />
compressed air for solvent and waste level sensing, and<br />
column/cartridge purging. Optionally, the system can deliver a<br />
compressed gas instead of air. Inert or other gases may be<br />
more suitable than air for some applications.<br />
If ambient air is compatible with your solvents and waste,<br />
simply leave the GAS INLET port open.<br />
If you want to use a gas instead of air, connect user-supplied<br />
1 8" I.D. tubing to the GAS INLET port. Connect the other end<br />
to the user-supplied compressed gas. The external compressed<br />
gas should be dry, filtered, and regulated to 2–5 psi (0.14–0.34<br />
atm, 13.8–34.5 kPa).<br />
2.9 Connect and Route Drain Line<br />
WARNING<br />
Risk of fire or equipment damage. Failure to connect drain lines<br />
may allow organic solvents to pool in unsafe areas, creating a<br />
potential for dangerous levels of flammable vapors. Improper<br />
draining may damage the instrument’s internal components.<br />
The Torrent has a front-mounted drip tray. There is a drain<br />
connection under the left side of the tray. Connect<br />
user-supplied tubing to this drain and route it to a suitable<br />
waste fluid collection container.<br />